ERROR CODES
During the initial power-on testing or during operation it is possible the scale may show
an error message. The meaning of the error messages is described below.
If an error message is shown repeat the procedure that caused the message, turning
the balance on, calibration or other functions. If the error message still is shown then
contact your dealer for further support.
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is
turned on or when the
ZERO/ENTER key is
pressed,
Weight on the pan when
turning the scale on.
Excessive weight on the
pan when zeroing the
scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Err 5
Keyboard Error.
Improper operation of the
scale.
Err 6
A/D count is not correct
when turning the scale on.
Platform not installed.
Load cell damaged.
Electronics damaged.
Err 8
Calibration Error
Attempt calibration again,
make sure the scale is
stable.
To view the A/D count press the ZERO/ENTER key and the % key at the same time
while the scale is performing the initial check at power-on. Press ZERO/ENTER to
return to normal.