11
Operating Instructions Chapter 3
English
Operating Instructions
3
!
C
General Instructions
1. Place the Inogen One® G3 in a well ventilated location.
Air intake and exhaust must have clear access.
Locate the Inogen One® G3 in such a way that any
auditory alarms may be heard.
WARNING
Avoid use of the Inogen One® G3 in presence of pollutants, smoke
or fumes. Do not use the Inogen One® G3 in presence of flammable
anesthetics, cleaning agents or other chemical vapors.
CAUTION
Do not obstruct air intake or exhaust when operating the device.
Blockage of air circulation or proximity to a heat source may lead to
internal heat buildup and shutdown or damage to the concentrator.
CAUTION
The Inogen One® G3 Concentrator is designed for continuous use. For
optimal sieve bed (columns) life, the product should be used frequently.
2. Ensure both particle filters are in place.
CAUTION
Do not operate the Inogen One® G3 without both
particle filters in place. Particles drawn into the
system may damage the equipment.
Exhaust
Intake
96-06947-00-01AG3 User Manual Inogen One G3HF-Multilanguage.indb 11
5/25/2016 4:47:00 PM
Summary of Contents for One G3
Page 2: ...2 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 2 5 25 2016 4 46 52 PM ...
Page 4: ...4 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 4 5 25 2016 4 46 52 PM ...
Page 14: ...10 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 10 5 25 2016 4 46 59 PM ...
Page 48: ...44 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 44 5 25 2016 4 47 52 PM ...
Page 50: ...46 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 46 5 25 2016 4 47 53 PM ...
Page 60: ...56 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 56 5 25 2016 4 52 35 PM ...
Page 94: ...90 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 90 5 25 2016 4 58 30 PM ...
Page 96: ...92 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 92 5 25 2016 4 58 30 PM ...
Page 106: ...102 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 102 5 25 2016 4 58 37 PM ...
Page 140: ...136 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 136 5 25 2016 4 58 46 PM ...
Page 142: ...138 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 138 5 25 2016 4 58 46 PM ...
Page 152: ...148 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 148 5 25 2016 4 58 51 PM ...
Page 186: ...182 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 182 5 25 2016 4 58 59 PM ...
Page 188: ...184 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 184 5 25 2016 4 58 59 PM ...
Page 198: ...194 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 194 5 25 2016 4 59 05 PM ...
Page 232: ...228 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 228 5 25 2016 4 59 13 PM ...
Page 234: ...230 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 230 5 25 2016 4 59 13 PM ...
Page 244: ...240 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 240 5 25 2016 4 59 18 PM ...
Page 278: ...274 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 274 5 25 2016 4 59 27 PM ...
Page 280: ...276 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 276 5 25 2016 4 59 27 PM ...
Page 290: ...286 96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 286 5 25 2016 4 59 32 PM ...
Page 323: ...96 06947 00 01AG3 User Manual Inogen One G3HF Multilanguage indb 319 5 25 2016 4 59 40 PM ...