Service Item
Estimated Service Requirement Frequency
Battery
500 full charge/discharge cycles (to approximately 80% capacity)
Compressor
Function of flow setting and time
17,000 run hours @ flow setting 5
20,000 run hours @ flow setting 2
Oxygen Sensor
No Service Requirement
Molecular Sieve
36 to 60 months depending on usage
Valves 20,000 run
hours
Intake Filter
20,000 run hours
Cooling Fan
20,000 run hours
4.
INOGEN ONE G2 SYSTEM SPECIFICATIONS
INOGEN ONE G2 CONCENTRATOR
Dimensions:
Length
Width
Height
Cm In Cm In Cm In
No battery
27.2 10.7 9.9 3.9
22.2 8.75
12 cell
battery
27.2 10.7 9.9 3.9
24.1 9.50
24 cell
battery
27.2 10.7 9.9 3.9
26.0 10.25
In Bag
27.2 10.7 11.4 4.50 25.4 10.00
On Cart
34.3 13.5 20.3 8.00 40.6 16.00
Weight:
Approximately 7.25 pounds (includes 12 cell battery)
Noise:
<38 dB on flow setting 2
Warm-Up Time:
Approximately 2 minutes
Oxygen Concentration: 90 + 6/-3%
Flow Control Settings:
1,2,3,4,5
Power:
AC/DC Power Supply
AC Input:
100 to 240 VAC, 50 to 60 Hz, 1.0A to 0.5A
DC Input: 13.5 to 15 VDC, 10A
Output 19 VDC, 90 W