3.2.3.
Surface Cleaning
The outside case should be cleaned using a cloth dampened with a solution of mild
detergent (e.g., dishwashing detergent, such as Dawn
TM
) and water or disinfecting
towelettes (e.g., DisCide Ultra).
3.2.4.
Battery Care and Maintenance
The Inogen One G2 Lithium Ion Battery requires special care to ensure proper
performance and long life. Use only Inogen One G2 Batteries with the Inogen One G2
Oxygen Concentrator.
DESIGN
NOTE
The Inogen One G2 Oxygen Concentrator adjusts its oxygen production rate to
match the oxygen demand specified by the user flow setting. When the device
is used at lower settings, its battery life is extended. Additionally, at lower
flow settings, the concentrator does not generate as much heat and noise,
draw as much electric current from external power supplies, and many system
components do not wear as quickly.
Initial Charging and subsequent charging
The Inogen One G2 battery requires an initial charge to full capacity before use.
Instruct the patient to avoid running the Inogen One G2 on battery power until
this initial charging has been completed. This charge typically takes 2 hours
depending on the flow setting.
The Inogen One G2 may be run on AC or DC power and used by the patient
during this initial charging period
Normal Charging
The battery is operating properly when a battery icon is displayed on the LCD
screen.
AMSR
oxygenconcentratorstore.com