INOGEN ONE G2 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 1 of 20
96-02735-00-01
Revision C
AMSR
oxygenconcentratorstore.com
1-877-774-9271
Page 1: ...ROPRIETARY 96 02735 00 01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL Page 1 of 20 INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL 96 02735 00 01 Revision C AMSR oxygenconcentratorstore com 1 877 774 9271 ...
Page 2: ...RATOR TECHNICAL MANUAL Page 2 of 20 This Manual applies to the following Inogen Inc products Inogen One G2 Oxygen Concentrator model IO 200 Inogen Universal Power Supply model BA 107 BA 207 Battery model BA 200 Battery model BA 224 AMSR oxygenconcentratorstore com 1 877 774 9271 ...
Page 3: ...ENT ON INOGEN ONE 6 2 1 RECOMMENDATIONS FOR USE 6 2 2 SYSTEM COMPONENTS 6 2 3 USING THE INOGEN ONE G2 7 2 4 SELECTING THE PROPER FLOW SETTING 7 3 SERVICING THE INOGEN ONE G2 9 3 1 MAINTENANCE BY THE PROVIDER 9 3 2 MAINTENANCE BY THE PATIENT 13 3 3 EXPECTED SERVICE REQUIREMENTS 16 4 INOGEN ONE G2 SYSTEM SPECIFICATIONS 17 5 INOGEN ONE G2 ERROR CODE TABLE 18 6 CONTACTS FOR MORE INFORMATION 20 AMSR ox...
Page 4: ...ntended to help assure that patients are given proper guidance in the use and function of the Inogen One G2 and its accessories Proper care in relaying this information will not only enhance the user s experience with the Inogen One G2 but will also protect the patient prolong the life of the device and help you avoid unnecessary service calls and complaints from users 1 1 Caution and Warning Stat...
Page 5: ...GN NOTE A DESIGN NOTE indicates specific information regarding the design of the Inogen One G2 and or accessories This information is included in this manual to provide you with a greater working understanding of the device This information is not required to operate or maintain the Inogen One G2 The Inogen One G2 User Manual contains additional cautions and warnings AMSR oxygenconcentratorstore c...
Page 6: ...al certifying bodies for Home Health Care Providers require that back up oxygen be available to the patient Supplemental oxygen cylinders or extra Inogen One G2 Batteries may satisfy these requirements CAUTION Oxygen demand of some patients particularly those with high breathing rates and high flow settings may exceed the capabilities of the Inogen One G2 Inogen suggests that each patient be titra...
Page 7: ...Bolus Volumes Specification All oxygen conserving devices OCD s function differently and therefore it is prudent to titrate patients for any new conserving device Delivery timing bolus volume and oxygen concentration all contribute to a patient s fraction of inspired oxygen FiO2 and therefore to the OCD s efficacy at maintaining the patient s blood oxygen saturation As an oxygen concentrator the I...
Page 8: ... 0 720 64 8 72 0 79 2 32 4 36 0 46 6 25 9 28 8 31 7 5 0 900 81 0 90 0 99 0 40 5 45 0 58 2 32 4 36 0 39 6 10 BPM ml 10 20 BPM ml 10 25 BPM ml 10 DESIGN NOTE Bolus Volume is tuned to provide the correct bolus volume when delivered through a nasal cannula such as the Salter Labs 1600Q If bolus volumes are measured without a cannula or with a different type of nasal cannula bolus volumes will vary fro...
Page 9: ...and SW version To check usage press and hold the Alarm Bell Button for 5 seconds The concentrator s display will show the hour meter the unit s serial number and the software version installed on the device NOTE Do not disassemble the Inogen One G2 or any of the accessories or attempt any maintenance other than tasks described in this Technical Manual unless you have completed a training course th...
Page 10: ...r cracks or other damage 4 Feel the sides of the device for vibration and listen for unusual noises rattles or other signs that the device requires service CAUTION Discovery of cracks or other types of external damage may be indicative of other internal damage that may not be visible If such external damage is discovered be certain to inquire as to how it occurred and whether any changes in the de...
Page 11: ...r 6 Carefully screw the nozzle fitting back into the recess clockwise until it bottoms out on the filter gasket Take care to squarely screw the nozzle fitting into the threads and not to over tighten WARNING Failure to inspect and replace the product filter may result in the filter becoming clogged or obstructed over time and in reduced delivery of oxygen to the patient 3 1 6 Data Logging Operatin...
Page 12: ... of the drive and the file system alone do not ensure compatibility with the Inogen One G2 Compatible USB drives include PNY Attache 2GB PN P FD2GBATT2 SF SanDisk Cruzer Micro 2GB PN SDCZ6 2048RB Kingston Data Traveler G3 2GB PN DTIG3 2GBZ Kingston DataTraveler 102 2GB PN DT102 2GB Kingston DataTraveler DT101 2GB PN DT101G2 2GB Dane Elec DNL 2GB PN DNLDAZMP02GCAW Transcend JetFlash V30 2GB PN TS2G...
Page 13: ...ith the Inogen One G2 unless proper saturation has been verified by a clinician 3 2 2 Intake Filter Cleaning At the front of the Inogen One G2 ambient air passes through a gross particle screen that removes dust fragments This particle screen must be cleaned on a weekly basis to ensure adequate air flow through the device If the particle screen is not cleaned frequently the life expectancy of the ...
Page 14: ...by the user flow setting When the device is used at lower settings its battery life is extended Additionally at lower flow settings the concentrator does not generate as much heat and noise draw as much electric current from external power supplies and many system components do not wear as quickly Initial Charging and subsequent charging The Inogen One G2 battery requires an initial charge to full...
Page 15: ...cally achieve 500 charge discharge cycles while retaining 80 of their original capacity if proper battery care is taken The BA 200 and BA 224 will typically achieve 500 charge discharge cycles while retaining 80 of their original capacity if proper battery care is taken Instruct the user that attempting to charge their battery while attached to a running Inogen One G2 concentrator at elevated temp...
Page 16: ...CAUTION Leaving a Battery in the Inogen One G2 while the device is unused will irreversibly damage the battery After such storage the battery may not be able to recharge or its life cycle and or capacity will be greatly diminished A full battery can be damaged in as little as 20 days if left in the concentrator while not plugged in Instruct users to avoid storing the Inogen One G2 battery in extre...
Page 17: ...ng Fan 20 000 run hours 4 INOGEN ONE G2 SYSTEM SPECIFICATIONS INOGEN ONE G2 CONCENTRATOR Dimensions Length Width Height Cm In Cm In Cm In No battery 27 2 10 7 9 9 3 9 22 2 8 75 12 cell battery 27 2 10 7 9 9 3 9 24 1 9 50 24 cell battery 27 2 10 7 9 9 3 9 26 0 10 25 In Bag 27 2 10 7 11 4 4 50 25 4 10 00 On Cart 34 3 13 5 20 3 8 00 40 6 16 00 Weight Approximately 7 25 pounds includes 12 cell battery...
Page 18: ...Type BF Not intended for cardiac application Degree of Protection Against Ingress of Water Concentrator IP 20 Carry bag IP 22 Degree of Safety for Application in Presence of Anesthetic Gases Not suitable for such applications 5 INOGEN ONE G2 ERROR CODE TABLE If more than one alarm is triggered simultaneously the error code number will be the sum of the error codes given below Error Code Number Exp...
Page 19: ...trator from the carry bag and restart the concentrator after 10 minutes to cool down If the error persists the fan may have failed If the System Cold message appears allow the concentrator to warm up for 15 minutes in a room temperature environment and then restart the concentrator Battery Hot Warning Battery Hot Shutdown Battery over temperature during discharge The concentrator might need to be ...
Page 20: ...battery Service Needed The compressor has reached maximum speed The compressor is at its end of life or there is a leak in the concentrator External Power Low The input voltage is less than 17V Remove power to the concentrator and remove and reinsert the battery Try another external power source such as DC power 6 CONTACTS FOR MORE INFORMATION Inogen Website Patient Issues http www inogenone com C...