oxygen is
rate-limited
. The Inogen At Home delivers up to 5 L/min of 90% (+6%/-3%)
oxygen.
At each flow setting, the Inogen One generates a specific amount of oxygen (1 L/min per
setting), which is delivered continuously to the patient. The device will deliver the same
volume of oxygen regardless of the patient’s breath rate.
Flow Setting
Flow Rate (L/min)
1
1.0
2
2.0
3
3.0
4
4.0
5
5.0
3. SERVICING THE INOGEN AT HOME
3.1. Maintenance by the Provider
3.1.1. Authorized Repair Centers
The Inogen At Home is only intended to be repaired by authorized repair centers. If a
repair is required, please contact Inogen to locate your nearest authorized repair
center. If you would like to become an authorized repair center, please contact
Inogen for access to certification training, service instructions, component part lists
and the necessary repair equipment.
3.1.2. Cleaning and Processing for Reuse
Reprocessing between patient use is required. Refer to the following guidance on
the process to clean and disinfect for reuse.
The following supplies are recommended:
o
Nitrile Gloves or equivalent
o
Safety Glasses
o
Disinfecting wipes or Spray designed to kill bacteria and viruses
o
Compressed Air (30 psi / 207 kPa, for blowing off parts)
o
Particulate Respirator (N95 rated)
The following safety precautions should be followed when cleaning and processing a
concentrator for reuse:
1. The cleaning process should be performed in a well-ventilated area.
2. Wear nitrile gloves, safety glasses, and a particulate respirator when
cleaning and disinfecting the concentrator.