innovu health INH-GB-01 User Manual Download Page 3

Shape Structure

1   

Dual mode with removable probe cover

2

   Infrared probe

3

   Measuring key

4

   LCD colour display

5   

Memory key and conversion key

6

   Handle

7

   Battery compartment cover

Principle

Any object in which temperature is higher than absolute zero degree will transmit some infrared 
radiation energy according to its own temperature. The radiation energy and its distribution per 
wavelength are closely associated with its surface temperature. Based on the principle, it is possible 
to measure the forehead surface temperature and then adjust the offset between forehead 
temperature and actual body temperature, which will result in the correct display of body 
temperature.

Product Features

Technical Parameters

Display Icon Definition

4

3

Contact type infrared measurement

3 modes of measurement; forehead, ear and physical object

High definition dual colour LED screen

Fever alarm function (default alarm value is 38.0°C and above)

Capacity to store 20 readings in memory

Optional unit of temperature; °C or °F

Automatic shut down and power saving

Measurement Method

Icon Definition

Measuring mode 1

Measuring mode 2

Temperature unit

Object temperature mode

Battery capacity

Icon

Icon Definition

Forehead and object temperature measurement

Ear temperature measurement

Celsius or Fahrenheit

The icon means you can convert between °F and °C

Object temperature measurement

Indicates low battery status

Contact Type
Body Mode

32.0°C - 42.9°C (89.6°F - 109.2°F)
0°C - 100°C (32°F - 212°F)
±0.2°C (0.4°F)
±1.0°C (1.8°F)

Surface Mode
Body Mode
Surface Mode
0.1°C/°F

16°C - 35°C (60.8°F - 95°F) <85% moisture condensation

25°C - 55°C (-13°F - 131°F) <90% moisture condensation

DC 1.5V x2
When off <10 uw
When measuring <30 mw
In 30 seconds
About 72g (without battery)
152mm x 44mm x 30mm (l x w x h)
This product complies with EMC standards

Measurement Mode

Measurement Accuracy

Display Resolution

Operating Environment

Storage Condition

Power Supply
Power Consumption

Auto Power-off

Weight
Dimensions
EMC

2

3

4

5

6

7

1

Summary of Contents for INH-GB-01

Page 1: ... HEALTH NON CONTACT INFRARED THERMOMETER INH GB 01 USER MANUAL innovuhealth co ...

Page 2: ...ccuracy with age so we advise replacing it Product Intended Use The infrared thermometer is intended for the measurement and monitoring of human body temperate for people of all ages Attentions Notes Before use please read the instructions and make sure the batteries provided have been installed Ensure this device does not have contact with liquid or extreme high low temperature Avoid this device ...

Page 3: ...unction default alarm value is 38 0 C and above Capacity to store 20 readings in memory Optional unit of temperature C or F Automatic shut down and power saving Measurement Method Icon Definition Measuring mode 1 Measuring mode 2 Temperature unit Object temperature mode Battery capacity Icon Icon Definition Forehead and object temperature measurement Ear temperature measurement Celsius or Fahrenhe...

Page 4: ...e MEM button and hold for 6 seconds and the screen will then display C If you want to change it to F press the MEM button once more and hold for 3 seconds C Voice switch setting Whenpoweredon holdtheMEMbuttonfor4seconds itwillbeeponceandthencloseoropenup Note thesoundkeycanbeseton offdependingontheenvironment e g itcanbeoffwhensilenceisappreciated D Shutdownkey Whenpoweredon holdtheMEMbuttonfor8se...

Page 5: ...The thermometer can then begin taking object measurements by pointing the device at the object at 3 5mm away Press and hold the MEM button When you hear a beep the temperature will then display on the screen The device only needs routine cleaning and period or times can be decided according to user s need The surface cleaning use a clean soft cloth or cotton swab stick moistened with a little medi...

Page 6: ...ains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of this device requires continued power without interruption we recommend this device is powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typi...

Page 7: ...nt should be used no closer to any part of the device than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P NOTE 1 At 90MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection fr...

Reviews: