Innovative instruments DMM B18 Manual Download Page 3

 

 

 

 

DMM B18  Manual  

(English)

 

 

 

 

 

 

JULY 2022 

Page

 

3

 

 

1.

 

INTRODUCTION 

 
1.1

 

Principle:  

DMM B18 digital moisture meter works on the radio-frequency (RF) dielectric method. This method 
measures moisture content in sample by sensing the dielectric constant of the sample. The dielectric constant 
is a measure of a material’s ability to store electrical charge when placed in an electric field. Because of its 
molecular structure, water has a very high dielectric constant (approximately 80) compared to other grain 
constituents (2 to 3). The RF dielectric method is influenced significantly by grain kernel structure and 
composition, moisture distribution within kernels and grain temperature, necessitating individual calibration 
equations for different grain types.  

1.2

 

Calibration method:  

DMM B18 digital moisture meter is calibrated using a standard process. This calibration process consists of 
determining individual calibration equation by establishing a co-relation between the measured dielectric 
constant of the sample and the standard reference method. The standard reference methods used are Oven 
Drying Method and Toluene method (Dean-Stark) as applicable for the sample. The calibration equation 
compensates the temperature variation of the sample too. 

I.

 

The Oven Drying Method: References for the standard methods used: 

 

 

ISO 712: 1998 : Cereal and cereal products :Determination of moisture content - Routine reference 
method 

 

 

IS 4333(part 2) : 2002 (Reaffirmed in 2012) : Method of analysis for food grains, Determination of 
moisture content 

Method

: The sample is ground after pre conditioning. It is dried in a preheated oven at a temperature 

of 130 

0

C+/- 3 

0

C for 2-3 hrs till constant weight of the sample is achieved. 

 

 

ISO 6540 : Maize-Determination of moisture content in milled grains and whole grains  

Method

: The sample is ground after pre conditioning. It is dried in a preheated oven at a temperature 

of 130 

0

C+/- 3 

0

C for 4 hrs till constant weight of the sample is achieved. 

 

 

ISO 665: 2002 - Determination of moisture and volatile matter for oilseeds 

 

 

IS 3579-1966 (Reaffirmed in 2000): Method of test for Oilseeds. 

Method

: The sample is ground after pre conditioning. It is dried in a preheated oven at a temperature 

of 103 

0

C+/- 2 

0

C for 6-7 hrs till constant weight of the sample is achieved. 

 

II.

 

Distillation Method (The Dean and Stark) : References for the standard methods used: 

 

 

ISO 939-2021: Spices and condiments- Determination of moisture content 

Method

: A known weight of sample is placed in a flask with an organic solvent toluene which is 

insoluble with water and has higher boiling point and less density than water. This flask is attached to 
a condenser and the mixture is heated. The water in the sample evaporates and moves up into the 
condenser where it is cooled and converted back into liquid water, which then trickles into the 
graduated tube. When no more water is collected in the graduated tube (this process takes around 1 
hr), distillation is stopped and the volume of water is read from the tube. This water is directly 
proportional to the moisture content in the sample. 

 

 

Summary of Contents for DMM B18

Page 1: ...MAHAJAN SOCIETY B H CONVENT SCHOOL FATEHGUNJ BARODA 390 002 GUJARAT INDIA Contact 91 265 2791184 91 6356615024 Or Email us at info innovative instruments in innovativeinstrumentsbaroda gmail com Visit...

Page 2: ...ION Page 3 2 MODELS Page 4 3 FEATURES Page 4 4 TECHNICAL SPECIFICATIONS Page 4 5 OPERATING PROCEDURE Page 5 6 PROCEDURE TO REPLACE BATTERIES Page 6 7 SAMPLE PREPARATION AND HANDLING Page 6 8 WARRANTY...

Page 3: ...eaffirmed in 2012 Method of analysis for food grains Determination of moisture content Method The sample is ground after pre conditioning It is dried in a preheated oven at a temperature of 130 0 C 3...

Page 4: ...asy to read display in Auto switch off after 10 minutes of no operation Automatic Internal Temperature Compensation Non destructive testing Easy to use Pour and get result Repeatable and accurate resu...

Page 5: ...e the empty hopper and keep aside viii Press the appropriate channel switch to start the moisture measurement will blink for few seconds then moisture will be displayed DO NOT SHAKE OR TOUCH THE METER...

Page 6: ...ed to warm to room temperature before testing Similar care should be taken for heated samples also Sample with water added externally will show higher moisture than the actual when tested immediately...

Page 7: ...l 9 0 Moth dal 9 5 Kabuli chana dal 10 0 Masoor dal 9 5 Moong dal chilka 10 0 Tivda dal 9 0 Urad dal 9 5 Peas dal 9 5 Cowpeas dal 10 0 OILSEEDS Groundnut 5 0 Mustard 6 5 Rui 7 0 Groundnut HPS 5 0 Kard...

Page 8: ...S Cashew Kernel 3 5 Almond 5 0 Raisin 11 0 Pista 3 5 Walnut 6 5 MEDICINAL Gloriasa Superba Herbal Seed 10 0 Isabgol 8 0 Isabgol Husk 10 0 Paracetamol Powder 8 0 MISCELLANEOUS Puvad Cassia seed 7 5 Guv...

Reviews: