4.

Other Commands for Controlling the ALPHAiX 3000

PERIOD

 key

The PERIOD key is used to program the measuring intervals.The currently set 
measuring time appears if PERIOD is pressed immediately after switching on. 
This measuring time display flashes if the SET key is depressed at the same 
time.The measuring time setting can be changed with the 

/

 keys in this 

phase. After the required measuring time is set, the SET key is depressed 
again until the time is displayed. The measuring range selection mode is then 
exited and other functions can be used.

If the PERIOD key is depressed during a measurement
-

the measuring time already elapsed is displayed first

-

on the 2nd press the total measuring time previously set is displayed 

-

on the 3rd press the unit returns to display the current measuring 

value.

The measured value is always shown in the display at the end of a series of 
measurements.

No total measuring time display is given when measuring 

µ

Sv/h because the 

measuring time is a variable in this measuring method - i.e. sometimes 
shorter, sometimes longer depending on the impulse density.

STORE

 key

Display of the average (mean value) from all stored (max. 43) measured 
values ( /0). This value (total measuring time should be at least 100 minutes) 
can be used as background rate if no other radiation exposure is present. 

After depression of STORE, pages of the stored values can be turned with the 

/

 keys, starting with the display of the date, time, background counts 

setting and measuring time. 

Reading the memory is, however, only possible if no measurement series is in 
progress - i.e. the measurement series must first be interrupted with STOP or 
the memory must be made accessible with STORE immediately after switching 
on. 

The measured values are numbered consecutively with the oldest value as No. 
1 and the latest measured value as No. 40.

SOUND

 key

The SOUND key allows to switch on/off the audible impulse signal in the 
IMPULSE mode. With this key the alarm can also be switched off. 

Simultaneous depression of the SOUND and SET key permits setting of the 
zero rate, ie. adjusting - with 

/

  keys  - the background count.

For dose rate measurements in 

µ

Sv it is important to first program the correct 

zero rate. A zero rate of 20 IPM/m is set as a factory default - which 
corresponds to the zero rate for the counter tube type A.

The zero rate for counter tube type B is 

8 IMP/m

The zero rate for counter tube type FSZ is 

17 IMP/m

The zero rate for counter tube type G is 

28 IMP/m

OPERATING INSTRUCTIONS 

ALPHAiX 3000

JS Friday 22 April 2011                                                                                                                                                     Page 5 of 7

Summary of Contents for ALPHAiX 3000

Page 1: ...ng intervals Page 4 4 Other commands for controlling the ALPHAiX Page 5 5 Replacing the counter tubes Page 6 6 Replacing the battery Page 6 7 Guarantee Page 7 8 EDP measurement analysis program for th...

Page 2: ...for approx 3 sec STOP appears on the LCD measurement is aborted can be re started by RUN STOP d Changing the measuring method 3 different measuring methods can be set by depressing the SET key Single...

Page 3: ...at measured value No 41 is stored as No 40 With this measuring method a shorter measuring interval should be chosen The mean value per minute from all stored measured values can later be retrieved aga...

Page 4: ...SE node The dots in the display disappear and only a flashing dot indicates that the measurement is continuing in the background The display is reactivated by depressing the RUN STOP and PERIOD keys a...

Page 5: ...ed values 0 This value total measuring time should be at least 100 minutes can be used as background rate if no other radiation exposure is present After depression of STORE pages of the stored values...

Page 6: ...s probes Geiger M ller counter tubes can be attached To replace the counter tube the unit must be switched off or the tube must be attached before switching on The unit must be left for approx 1 2 min...

Page 7: ...different counter tubes for connection to the ALPHAiX 3000 are available Counter tube type A for measuring Alpha Beta and Gamma radiation This counter tube is especially suited for educational facilit...

Reviews: