Innova ALPHAiX 3000 Operating Instructions Manual Download Page 4

3.

Programming the Measuring Time/Measuring Interval

The ALPHAiX allows setting of various measuring times in the IMPULSE mode or 
measuring intervals for display in IMP/m but not in 

µ

Sv/h. As already mentioned, the 

new measuring intervals in the 

µ

Sv/h and IMP/m measuring method are 

automatically started after the pre-selected measuring time and the measured value 
of the completed measuring interval is stored. In other words, it concerns indefinite 
measurement series.

This is not possible in the IMPULSE mode. The unit stops after the set measuring 
time showing the total number of impulses counted during the measuring time in 
the display. The unit must be re-started manually (RUN key), which re-sets the 
display to zero.

For the IMPULSE mode a longer measuring time is usually selected than for the 
measuring methods (

µ

Sv/h & IMP/m) which involve automatic storage of the 

measured values. To allow retrieval of intermediate results in the IMPULSE mode the 
count can be stopped without interrupting it. 
The display stops if the RUN/STOP + PERIOD keys are depressed simultaneously in 
the IMPULSE node. The dots in the display disappear and only a flashing dot 
indicates that the measurement is continuing in the background. The display is 
reactivated by depressing the RUN/STOP and PERIOD keys again, i.e. the impulses 
counted during that time are loaded to the display and the measurement is 
continued with the current values.

Simultaneous depression of the SET + PERIOD keys and the 

/

 keys means 8 

different measuring times/intervals can be set:

IMPULSE mode: 

10 sec. 

Imp/m + 

µ

Sv/h

/

10 sec

1 min

 

1 min

10 min

10 min

60 min

60 min

100 min

100 min

6 hours

6 hours

12 hours

12 hours

indefinite

24 hours

The difference in the last stage follows from the fact that the measuring methods 
IMP/m + 

µ

SV/h run indefinitely anyway. The current flashing measuring time 

appears on depression of the SET + PERIOD keys. The required measuring time can 
be set with the 

/

 keys. After the required setting the SET key must be depressed 

until the date display appears. A new series of measurements can then be started.

The 

/

 keys are identical to the TIME + SOUND keys and are generally used to 

program the ALPHAiX. Programming is always completed with the SET key - i.e. it 
must be depressed until the date display appears and can be switched over.

OPERATING INSTRUCTIONS 

ALPHAiX 3000

JS Friday 22 April 2011                                                                                                                                                     Page 4 of 7

Summary of Contents for ALPHAiX 3000

Page 1: ...ng intervals Page 4 4 Other commands for controlling the ALPHAiX Page 5 5 Replacing the counter tubes Page 6 6 Replacing the battery Page 6 7 Guarantee Page 7 8 EDP measurement analysis program for th...

Page 2: ...for approx 3 sec STOP appears on the LCD measurement is aborted can be re started by RUN STOP d Changing the measuring method 3 different measuring methods can be set by depressing the SET key Single...

Page 3: ...at measured value No 41 is stored as No 40 With this measuring method a shorter measuring interval should be chosen The mean value per minute from all stored measured values can later be retrieved aga...

Page 4: ...SE node The dots in the display disappear and only a flashing dot indicates that the measurement is continuing in the background The display is reactivated by depressing the RUN STOP and PERIOD keys a...

Page 5: ...ed values 0 This value total measuring time should be at least 100 minutes can be used as background rate if no other radiation exposure is present After depression of STORE pages of the stored values...

Page 6: ...s probes Geiger M ller counter tubes can be attached To replace the counter tube the unit must be switched off or the tube must be attached before switching on The unit must be left for approx 1 2 min...

Page 7: ...different counter tubes for connection to the ALPHAiX 3000 are available Counter tube type A for measuring Alpha Beta and Gamma radiation This counter tube is especially suited for educational facilit...

Reviews: