20
M E N U 1
Use the
UP
or
DOWN
keys to select an action. Press
ENTER
to access the menu for that action. Press
EXIT
to return to the Standby screen.
E R A S E
D A T A
–
Select this option to erase data. Select either
Yes
or
No
to erase data and return to
the Standby screen.
F A C T O R Y
C A L
R E S E T
–
Select
Yes
to reset the factory calibration settings and automatically
return to the Standby screen or No to go directly back to Standby.
C A L I B R A T I O N
–
Once Calibration menu option has been selected, the analyzer will purge itself,
perform a zero, and then proceed to the next step. When zeroing is complete, you will be prompted to
Enter Tag Values. The tag values should be entered just as they are on the calibration bottle. You are not
required to calculate the PEF value for HC. Instead enter the propane value and the analyzer will
internally convert it to a hexane HC value. When entering tag values for CO and CO2 are entered are
percentages. For HC and NOx, values are entered as ppm. Any gas you do not wish to change the
calibration on should be set to ¯ ¯ ¯ which indicates no value change. O2 is self calibrating on atmospheric
O2 (air). To enter the tag values of your calibration gas, use the GAS key to select the gas. Press + to
change the value of the highlighted numeral and select the
NEXT
key to advance to the following
numeral. When finished, press Continue to advance to the next sequence serene. The screen will prompt
you to Turn Cal Gas On and all the values will be shown below. Once the cal gas has been turned on, wait
for the values to rise and stabilize then press Continue. This is when the analyzer performs its actual
calibration. Do not disturb the analyzer during this period or the accuracy of your calibration could be
jeopardized. When calibration is complete, you will be prompted to Turn Calibration Gas Off. When this is
completed, press
DONE
, the analyzer will turn on its pump, purge itself, perform an automated zero, and
return to the Standby screen. The analyzer is now ready to resume service.
S E R I A L P O R T
–
This option defines what is connected to the serial port. The two choices are
PC
(to connect to a PC or PDA) or
PRINTER
to connect to an external printer.