Multitron Pro - Incubation Shaker
Technical data
Page
46
of
185
2013-02-19
Mobile Pt100-Sensor
Description
Value
Control
Electronic PID-Controller
Sensor
Pt100
Control accuracy
0.2 °C
Measure accuracy
0.15 °C
UV-sterilization of process air
Abb. 7
Description
Value
UV-sterilization lamp 5 Watts socket G23
5 Watts
Wave length UV-C
200 – 280 nm
Maximum emission
253.7 nm
Life cycle
9000 h
Output UV-C radiation
1.1 Watts
Intensity after 5000 h
80%
Output total
11 Watts
Power consumption total
0.18 A
3.5 Operating conditions
Description
Value
Ambient temperature
max.
With / without cooling 30 °C
Ambient temperature
max.
without cooling:
8 °C below required minimum
incubation temperature
with cooling:
side cooling: 15 °C
top cooling: 12 °C
upon required minimum incubation
temperature
Relative humidity
max.
With / without cooling: 85% rH