Checking the system with control solution
Checking the system with control solution
• Glucose test
Testing procedure
1. Shake the control solution before use.
2. Discard the 1st drop of control solution, this will
eliminate any residue.
3. Place a drop of control solution on a clean, dry
surface(ex. The lid of the test strip vial).
4. Insert a test strip into the port firmly and press
Up(
▲
) button. The symbol of control solution
( ) will appear on the display of meter. This
will allow you to differentiate between a control
solution test and an actual blood test for future
reference.
5. Apply the test strip to the control solution.
3. Check the code number on the display of the meter with
that on the strip vial.
4. Insert the test strip and press Mode( ) button.
Press Up(
▲
) button and the meter displays the symbol of
control solution( ).
5. In 5 seconds the symbol of sample blood is blinking.
6. Place a drop of solution on a clean, dry
surface (for example: the lid of your test
strip vial).
7. Apply the prepared control solution to the
test strip by using capillary rod.
8. Apply the control solution to the test area.
Caution
: If you want to test all three test items (TC, HDL
and TG), please collect the control solution with
each capillary rod.
Caution
: Whenever you change the test strip vial,
Please follow the code recognition procedure.
9. The test results will be displayed in about 2 minutes.