Infopia GlucoLab Instructions For Use Manual Download Page 20

Trouble Shooting

MESSAGE

Problem                            Action Required

The ambient temperature              Place the meter within the 

is less than 10°C (50°F)                 operating temperature range 

for more than 30 minutes
and retest.

The ambient temperature is           Place the meter within the  

over 40°C (104°F)                               operating temperature range

for more than 30 minutes
and retest.

The test result is lower than 

10 mg/dl or 0.6 mmol/L

The test result is higher than 

600 mg/dl or 33.3 mmol/L

Check the accuracy of your test
strip by perfoming a control
solution test. With normal result,
re-test with blood sample two or
three times. When  “Lo” message
persists, please consult a doctor
for assistance.

Check the accuracy of your
test strip by perfoming a control
solution test. With normal
result, re-test with blood
sample two or three times.
When “HI” message persists,
please consult a doctor for
assistance.

38

39

MESSAGE

Problem                            Action Required

Problem with the test strip            Please insert a new test strip
Test strip is damaged.                  and perform your test again.

User applied blood sample           Please wait for the LCD to              

before the LCD display,                display,               blinking before  

blinking.                          applying your blood sample.

No readings in memory to display

Not enough readings in memory  
to display designated averages

Trouble Shooting

Summary of Contents for GlucoLab

Page 1: ...59 54 Fax 32 2 732 60 03 E Mail mail obelis net INFM08A IMC00099 Rev 2014 06 02 Lancet infopia Co Ltd Lancing Device GMMC Rm 1112 Ace Tower 9th Bldg 130 Digital ro Geumcheon gu Seoul Korea Instruction...

Page 2: ...Blood Glucose Monitoring System for self testing Quick and Smart Blood Glucose Monitoring System for self testing...

Page 3: ...is important to complete the warranty registration card included in your kit and mail it back to Obelis S A Doing this will help us serve your needs better Intended Purpose The Blood Glucose Monitorin...

Page 4: ...b Meter 34 Changing Your Battery 35 Disposal of meter and batteries 36 Trouble Shooting 37 40 Specifications 41 Warranty 42 Symbol Reference 43 7 6 Caution Before using any product please read all ins...

Page 5: ...attery Carrying case Your Blood Glucose Monitoring System has been sealed and it should contain all of the parts shown above Please return the system to place of purchase or contact your nearest repre...

Page 6: ...only DO NOT RE USE 8 Keep the blood glucose test strip vial away from children 9 Pleaserefertotheuser smanualfor bloodglucoseteststripforadditionalinformation 10 Discard the used blood glucose test s...

Page 7: ...Do not reuse Dehydration may lower test results If you are severely dehydrated contact you physician immediately Inaccurate results may occur when in shock hypotensive hyperglycemic or hyperosmolar st...

Page 8: ...eTemperatureUnit Settingtestingaverages Setting the Alarm After pressing and releasing either or buttons to set or enter After pressing and releasing either or buttons to set the average of testing da...

Page 9: ...ut be careful not to bend The code number should appear in the LCD window for 3 seconds example C18 Compare the code number on the meter s LCD window with the code number on the test strip vial If the...

Page 10: ...a convenient medical tool for collecting capillary blood samples for glucose monitoring or other tests which require one or two drops of blood As a more advanced lancing device it provides ultimate sa...

Page 11: ...port and re start the procedure from the beginning If the code on the LCD display window and on the test vial do not match follow the coding procedure for accurate results refer p18 If you apply your...

Page 12: ...he Hi Lo message will be shown on the LCD refer p39 Safely discard used blood glucose test strips and lancets Test Strip Insertion Completely filled Poorly filled Applying Blood Step 2 Place your fing...

Page 13: ...with your result then press Record the result value in your logbook When glucose test strip is removed glucose test meter turns off automatically CAUTION Although you already set the user option and i...

Page 14: ...ontrol solution should be used Whenever you suspect the blood glucose test meter or blood glucose test strip is not functioning properly If your blood glucose test results are not consistent with your...

Page 15: ...0 test results in its built in memory and records the average glucose level for the number of days which you had preset to be calculated and saved These values can be recalled in the LCD display at an...

Page 16: ...physical exercise or whenever you feel that your glucose levels may be changing rapidly should be done from the fingertip You should also use fingertip testing whenever you have a concern about hypog...

Page 17: ...into the channel of the blood glucose test strip You may get an inaccurate result if the blood sample is not completely filled in a single action To bring fresh blood to the surface of the test site r...

Page 18: ...he outside of the blood glucose test meter Your blood glucose test meter is a precision instrument Please handle it with care Lancet Device Clean the lancet device and caps with soap and tepid water T...

Page 19: ...and Council Directive 2012 19 EU aimed to reducing the amount of EEE waste going for final disposal The manufacturer has specific instructions for the recovery of the meter Please contact your distri...

Page 20: ...l result re test with blood sample two or three times When Lo message persists please consult a doctor for assistance Check the accuracy of your test strip by perfoming a control solution test With no...

Page 21: ...Shooting MESSAGE Problem Action Required The meter does not Battery is dead or there is a Change the battery and if power on problem with the meter the problem persists contact your local representat...

Page 22: ...acturer Authorized representative in the European Community Biohazard Waste batteries and accumulators 42 43 Warranty Five Year Warranty The meter has a full five year warranty from the original date...

Page 23: ...Memo 44 INFM08A IMC00099 Rev 2014 06 02...

Reviews: