INFINITY CLASSIA PSW310W (120V)
1. Select a new test position: Cue the test track corresponding to
the center frequency of the dip. In the example in Figure 19
you would play Tracks 14 (52Hz) and 18 (40Hz). Press
Play
›
.
You will see a reading very close to what you had before. Now,
slowly move the RSLM around the area, if possible remaining
within about a foot of the original test point. As you move the
RSLM, watch the bar graph. You will observe large level
fluctuations. Find a location for the subwoofer or a test
location that raises the response at these frequencies. You may
find it helpful to move the RSLM vertically. Dips can be
oriented in any axis.The position that restores the level to
about that of the adjacent test points is your new test position.
2. Reset the test level: Return to the section “Setting the
Subwoofer Test Level” on page 7. Perform the procedure
as described.
3. Repeat the measurements: Now that you are familiar with the
measurement process, you can go much faster by using Tracks
27–50. These tracks contain all the test tones necessary for
measurement. However, each test is only about three seconds,
and there is no frequency announcement. The first test is
100Hz. Just place each test mark in order until finished.
Connect the dots.
Your second measurement will no longer exhibit the deep
response dips. However, the peak will still be evident. Without
the influence of the response dips, the amplitude and center
of the peak may have changed.
4. Interpret the new data: Compare your new data to the examples
given in this section of the manual. Follow the instructions for
the example that most closely matches your new measurement.
Example 6.
Ideal Response:
If your test data looks similar to the example in Figure 21, you
have a very favorable setup. Skip to the “Final System Balance”
section, page 15.
Adjusting the R.A.B.O.S. Equalizer
Now that you have performed the measurements and interpreted
the data, you have the information needed to adjust the
subwoofer’s equalizer.
There are three equalizer adjustments on the subwoofer. Left
to right, they are marked Frequency, Level and Width. Each
control has 21 positions.These are numbered from left to right.
Therefore, Position 1 is the full counterclockwise position. The
following table illustrates all switch positions.
Position
F (Hz)
L (dB)
W
1 CCW
20
–
14.1
4.5%
2
20
–
13.9
5%
3
20
–
13.5
7.5%
4
21
–
13.1
10%
5
22
–
12.7
12.5%
6
24
–
11.7
16.5%
7
26
–
11.0
20.5%
8
30
–
10.2
23%
9
35
–
9.5
26%
10
38
–
8.9
28%
11
40
–
8.3
29.5%
12
43
–
7.9
31%
13
46
–
6.4
34%
14
49
–
4.4
39%
15
52
–
2.9
41.5%
16
56
–
1.9
43.5%
17
63
–
1.1
45%
18
66
–
0.5
46.5%
19
72
0.0
48%
20
77
0.0
49%
21 CW80
0.0
49.5%
Figure 20. Ideal response, no EQ needed
14