Standard Definitions 

Vector Network Analyzer hardware and test cables have a set of well understood systematic errors that 

affect the unprocessed measurements made by the instrument. The calibration standards in this kit have 

precisely-known electrical behavior, and during calibration the VNA software uses the raw measurement 

data and the known behavior of the standards to calculate the phase and magnitude of up to 12 complex 

error terms at each frequency point of the calibration. Once calibrated, the instrument applies Vector Error 

Correction to each data point measured.   

 

 

Summary of Contents for Fairview Microwave FMCK1025

Page 1: ...Electronics Inc User Manual N Coaxial Calibration Kit DC to 18 GHz Models FMCK1025 FMCK1026 Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 Tel 1 800 715 4396 or 972 649 6678 Fax 972 64...

Page 2: ...t be returned to Fairview Microwave and must be issued a return authorization number by Fairview Microwave prior to shipment Buyer shall prepay shipping charges to Fairview Microwave Obligation is lim...

Page 3: ...nty 2 General Information 4 Calibration Kit Description 4 Maintenance 4 Calibration 4 Supporting Test Port Adapters 4 Electrostatic Discharge Precautions 4 Connector Description 5 Connector Care 5 Con...

Page 4: ...pection and cleaning of mating surfaces is recommended Failure to do so may result in degraded repeatability and accuracy as well as damage to any mated devices Calibration To maintain and certify the...

Page 5: ...er connector contact pin depths should also be verified through regular inspections using a connector gage kit to ensure that connectors on both calibration devices and devices under test DUTs have co...

Page 6: ...stationary while torquing the nut on the other device or cable This open end wrench is supplied with the calibration kit for this purpose Using the torque wrench Hand tighten the connection being torq...

Page 7: ...ea Open male FMSC3039 1 ea Load female FMTR1068 1 ea Load male FMTR1069 1 ea 9 16 Open End Wrench FMTL1004 1 ea 13 16 Torque Wrench ST N 1316 BO14 1 ea Short female FMSC3023 1 ea Short male FMSC3024...

Page 8: ...ments made by the instrument The calibration standards in this kit have precisely known electrical behavior and during calibration the VNA software uses the raw measurement data and the known behavior...

Page 9: ...2020 Infinite Electronics Inc Fairview Microwave is a registered trademark of Infinite Electronics Inc 9...

Page 10: ...ttps www fairviewmicrowave com nsearch aspx Category Calibration Kits Portable Calibration Kits Calibration Kits Components keywords calibration ports vna sort y searchtype 1 view_type grid Fairview M...

Page 11: ...ics Inc 11 Contacts Customer Support Sales Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 USA Phone 800 715 4396 972 649 6678 Fax 972 649 6689 Sales Email sales fairviewmicrowave com Cu...

Reviews: