Infinite Fairview Microwave FMCK1017 User Manual Download Page 8

User Manual | 2.4mm Coaxial Calibration Kits 

 

 

8

 

   

 

 

 

 

 

 

 

        

    

fairviewmicrowave.com

 

 

Standard Definitions 

Vector Network Analyzer hardware and test cables have a set of well understood systematic errors that 
affect the unprocessed measurements made by the instrument. The calibration standards in this kit have 
precisely-known electrical behavior, and during calibration the VNA software uses the raw measurement 
data and the known behavior of the standards to calculate the phase and magnitude of up to 12 complex 
error terms at each frequency point of the calibration. Once calibrated, the instrument applies Vector Error 
Correction to each data point measured.   

 

Summary of Contents for Fairview Microwave FMCK1017

Page 1: ...ectronics Inc User Manual 2 92mm Coaxial Calibration Kit DC to 40 GHz Models FMCK1017 FMCK1018 Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 Tel 1 800 715 4396 or 972 649 6678 Fax 972...

Page 2: ...be returned to Fairview Microwave and must be issued a return authorization number by Fairview Microwave prior to shipment Buyer shall prepay shipping charges to Fairview Microwave Obligation is limi...

Page 3: ...nty 2 General Information 4 Calibration Kit Description 4 Maintenance 4 Calibration 4 Supporting Test Port Adapters 4 Electrostatic Discharge Precautions 4 Connector Description 5 Connector Care 5 Con...

Page 4: ...inspection and cleaning of mating surfaces is recommended Failure to do so may result in degraded repeatability and accuracy as well as damage to any mated devices Calibration To maintain and certify...

Page 5: ...aces should be visually inspected under magnification and cleaned on a regular basis Proper connector contact pin depths should also be verified through regular inspections using a connector gage kit...

Page 6: ...stationary while torquing the nut on the other device or cable This open end wrench is supplied with the calibration kit for this purpose Using the torque wrench Hand tighten the connection being torq...

Page 7: ...e FMTR1057 1 ea Load male FMTR1058 1 ea 1 4 X 5 16 Open End Wrench FMTL1001 1 ea 7 16 X 1 2 Wrench FMTL1002 1 ea 5 16 Torque Wrench ST SMA 516 BO8A 1 ea Short female FMSC3012 1 ea Short male FMSC3013...

Page 8: ...ments made by the instrument The calibration standards in this kit have precisely known electrical behavior and during calibration the VNA software uses the raw measurement data and the known behavior...

Page 9: ...2019 Infinite Electronics Inc Fairview Microwave is a registered trademark of Infinite Electronics Inc 9...

Page 10: ...ttps www fairviewmicrowave com nsearch aspx Category Calibration Kits Portable Calibration Kits Calibration Kits Components keywords calibration ports vna sort y searchtype 1 view_type grid Fairview M...

Page 11: ...ics Inc 11 Contacts Customer Support Sales Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 USA Phone 800 715 4396 972 649 6678 Fax 972 649 6689 Sales Email sales fairviewmicrowave com Cu...