User Manual | 2.4mm Coaxial Calibration Kits 

 

 

6

 

   

 

 

 

 

 

 

 

        

    

fairviewmicrowave.com

 

 

Connector Tightening 

Damage to a calibration standard or attaching connector can occur if the device is turned instead of the 
connector’s coupling nut. ALWAYS turn the coupling nut when making connections. Never turn or spin the 
connectors. 

Always use a torque wrench (Fairview Microwave  model  ST-SMA-516-BO8A) to final-tighten all 
connections. This will ensure calibration accuracy and measurement repeatability. 

When making connections, an open-end wrench is recommended to hold the body of one device stationary 
while torquing the nut on the other device or cable. This open-end wrench is supplied with the calibration 
kit for this purpose. 

Using the torque wrench: 

Hand-tighten the connection being torqued by holding the calibration device steady and turning only the 
coupling nut. 

 

Hold the torque wrench with your thumb and index finger, behind the groove in the handle (See 

Figure 1

.). 

 

Tighten the connection until the ball in the handle crests on the cam (as the handle begins to break 
over). Do not “fully break” the handle of the torque wrench to reach the specified torque. 

 

Reverse the previous procedure to disconnect.

 

 
 
 

 
 

 

 
 
 

Torque wrench handle will break over when preset torque is achieved. 

Take note of the wrench bit orientation relative to the direction of motion/rotation.

 

 

Summary of Contents for Fairview Microwave FMCK1017

Page 1: ...ectronics Inc User Manual 2 92mm Coaxial Calibration Kit DC to 40 GHz Models FMCK1017 FMCK1018 Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 Tel 1 800 715 4396 or 972 649 6678 Fax 972...

Page 2: ...be returned to Fairview Microwave and must be issued a return authorization number by Fairview Microwave prior to shipment Buyer shall prepay shipping charges to Fairview Microwave Obligation is limi...

Page 3: ...nty 2 General Information 4 Calibration Kit Description 4 Maintenance 4 Calibration 4 Supporting Test Port Adapters 4 Electrostatic Discharge Precautions 4 Connector Description 5 Connector Care 5 Con...

Page 4: ...inspection and cleaning of mating surfaces is recommended Failure to do so may result in degraded repeatability and accuracy as well as damage to any mated devices Calibration To maintain and certify...

Page 5: ...aces should be visually inspected under magnification and cleaned on a regular basis Proper connector contact pin depths should also be verified through regular inspections using a connector gage kit...

Page 6: ...stationary while torquing the nut on the other device or cable This open end wrench is supplied with the calibration kit for this purpose Using the torque wrench Hand tighten the connection being torq...

Page 7: ...e FMTR1057 1 ea Load male FMTR1058 1 ea 1 4 X 5 16 Open End Wrench FMTL1001 1 ea 7 16 X 1 2 Wrench FMTL1002 1 ea 5 16 Torque Wrench ST SMA 516 BO8A 1 ea Short female FMSC3012 1 ea Short male FMSC3013...

Page 8: ...ments made by the instrument The calibration standards in this kit have precisely known electrical behavior and during calibration the VNA software uses the raw measurement data and the known behavior...

Page 9: ...2019 Infinite Electronics Inc Fairview Microwave is a registered trademark of Infinite Electronics Inc 9...

Page 10: ...ttps www fairviewmicrowave com nsearch aspx Category Calibration Kits Portable Calibration Kits Calibration Kits Components keywords calibration ports vna sort y searchtype 1 view_type grid Fairview M...

Page 11: ...ics Inc 11 Contacts Customer Support Sales Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 USA Phone 800 715 4396 972 649 6678 Fax 972 649 6689 Sales Email sales fairviewmicrowave com Cu...

Reviews: