7 – 17
Transpector SPS Operating Manual
2
Measure the resistance of each of the pins with respect to ground
(pin
7
and
8
). These measurements must be above 30 M
.
3
Measure the resistance of each of the pins with respect to each other. All of
these measurements must be above 30 M
, with the exception of the
measurement between the filament pins if the filament has not failed.
7.7.2 Transpector Sensor Filament Replacement
1
Remove the ceramic sealing disk by carefully pulling the wire hold downs away
from the edges of the disk and pulling off the sealing disk.
2
Remove spring.
3
Remove the filament lead screw (A), loosen the filament lead screw (B) and
remove the filament leads from the filament. (See
Figure 7-17 Sensor head
Ceramic
Wire Hold
Spring
Screw
Filament Leads
Sealing
Disk
Downs
A
Screw
B