1.16.2 Electronics Module Installation
The Transpector MPS electronics module should be mounted in an area where the
ambient temperature does not exceed 50°C and there is free air circulation around
the electronics module. Best performance will be achieved if the electronics
module is not located close to major heat sources where it is subjected to wide
temperature variations. For
Electronics module dimensions, see section Figure 1-5
After the sensor has been installed on the vacuum system, the Transpector MPS
electronics module must be mounted on the sensor:
1
The Transpector MPS sensor mounting connector assembly includes a
mounting nut and an O-ring. When the mounting nut is tightened, the O-ring
compresses making a tight fit on the sensor housing. For proper installation,
place the nut over the end of the sensor and roll the O-ring back to the groove
on the sensor.
2
Note the recessed area on the sensor feedthrough and the ground tab on
Transpector MPS electronics module. Match the recessed area of the
feedthrough to the ground tab and carefully slide the Transpector MPS
electronics module onto the sensor. Make sure the Transpector MPS
electronics module slides on fully.
3
Hand tighten the mounting nut on the Transpector MPS sensor.
4
and install the communications cable.
Figure 1-5 Electronics module dimensions
121.67 mm
60.32 mm
57.48 mm
114.96 mm
(4.79 in.)
(2.375 in.)
(4.526 in.)
(2.263 in.)