1 - 7
Transpector MPS Operating Manual
1.9.3 Mounting Requirements
The sensor is mounted to a high-vacuum chamber with a 69.9 (2.75 in.) DN40
ConFlat
®
flange with an extension tube with an OD of 41.2 mm (1.62 in.) and a
38.1 mm (1.5 in.) ID.
The electronics module attaches to and is supported by the sensor.
Transpector MPS can be mounted in any position.
1.9.4 Ventilation Requirements
At least 25.4 (1 in.) of open space around the Transpector MPS electronics module
must be maintained for proper ventilation.
1.9.5 Maintenance Access
Easy access to Transpector MPS should be maintained for installation and
maintenance activities.
1.10 Electrical Power Requirements
Transpector MPS must be connected to a source of power as specified in the
following sections.
1.10.1 Required Supply Voltage
20 to 30 V(dc), 24 V(dc) typical
1.10.2 Current Rating
1.25 A maximum
1.10.3 Electrical Connection
Latching, 4-pin DIN connector, internally isolated from system ground.
See
Figure 1-2 Transpector MPS connections
Electrical Connection