Equipment Description . . . . . . . . . . . . . . TEK-Mate® Refrigerant Leak Detector
Applicable Directives . . . . . . . . . . . . . . . 2006/95/EC(LVD)
2004/108/EC(EMC)
2011/65/EU(RoHS)
Applicable Standards . . . . . . . . . . . . . . . EN61010-1: 2010 EN55011, Group 1,
Class A: 2010 EN 61000-6-1:2007
EN14624:2012
CE Implementation Date. . . . . . . . . . . . . Sept 8, 2014
Authorized Representative . . . . . . . . . . . Brian King
Business Line Manager
Service Tools
Any questions relative to this declaration or to the safety of INFICON's products should be
directed, in writing to the quality assurance department at the above address.
WARNING
This symbol is intended to alert the user to the presence of important
operating and maintenance (servicing) instructions in the literature
accompanying the instrument.
TEK-Mate®, Toolbox Tough™ and INFICON® are trademarks of INFICON Inc.
DURACELL® is a registered trademark of Duracell Inc.
Specification Table in Accordance with EN 14624
Minimum sensitivity to R134a, fixed (static)
2 g/yr
Maximum sensitivity to R134a, fixed (static)
>50 g/yr
Minimum sensitivity to R134a, moving (dynamic)
2 g/yr
Maximum sensitivity to R134a, moving (dynamic)
>50 g/yr
Minimum response/detection time
<1 second
Zeroing time
5-7 seconds
Recovery time for 50 g/yr exposure*
12 second
Minimum sensitivity in contaminated environment
2 g/yr
Calibration frequency: Check annually with calibrated leak standard.
*Upper leak detection limit is not specified by INFICON as there is no upper limit to the size
of the leak the detector is able to detect. As no 50 g/yr leak standard was available during
testing, a 31 g/yr leak was substituted.