12 - 1
IP
N 07
4-
50
5-
P1
E
IC6 Operating Manual
Chapter 12
Maintenance and Calibration Procedures
12.1 Importance of Density, Tooling and Z-Ratio
The quartz crystal microbalance is capable of precisely measuring the mass added
to the face of the oscillating quartz crystal sensor. The IC6's knowledge of the
density of this added material (specified in the density parameter in Material
Set-Up) allows conversion of the mass information into thickness. In some
instances, where highest accuracy is required, it is necessary to make a density
calibration as outlined in
.
Because the flow of material from a deposition is not uniform, it is necessary to
account for the different amount of material flow onto the sensor compared to the
substrates. This factor is accounted for in the tooling parameter in Material Set-Up.
The tooling factor can be experimentally established by following the guidelines in
In the IC6 if the Z-Ratio is not known it could be estimated from the procedures
outlined in
, or, typically, the Auto-Z function can be used
to determine the Z-Ratio.
12.2 Determining Density
NOTE:
The bulk density values retrieved from the Material Library are sufficiently
accurate for most applications.
Follow the steps below to determine density value.
1
Place a substrate (with proper masking for film thickness measurement)
adjacent to the sensor, so that the same thickness will be accumulated on the
crystal and substrate.
2
Set density to the bulk value of the film material or to an approximate value.
3
Set Z-Ratio to 1.000 and tooling to 100%.
4
Place a new crystal in the sensor and make a short deposition
(1000-5000 Å) using manual control.
5
After deposition, remove the test substrate and measure the film thickness with
either a multiple beam interferometer or a stylus-type profilometer.
Summary of Contents for IC6
Page 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Page 2: ......
Page 8: ......
Page 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Page 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...