7 Warning and error messages
During operation, the display shows information that helps you operate the device.
Measurement values are displayed along with current device modes, operating
instructions as well as warnings and error messages.
The device is equipped with extensive self-diagnostic functions. If the electronics
detect a faulty state, the device will show this as far as possible on the display and will
interrupt operation when necessary.
Warnings
Warnings warn of device states that can impair the accuracy of measurements.
Operation of the instrument is not interrupted.
Pressing the "X" button acknowledges the warning.
Error messages
Errors are events that the device cannot correct itself and that force interruption of its
operation. The error message consists of a number and a descriptive text.
If you have eliminated the cause of the error, then start operation again by pressing
the button
.
The following table displays all the warnings and error messages. It states possible
causes for the malfunction and instructions on troubleshooting.
No.
Message
Possible error sources
Troubleshooting
1xx system error
W102
Timeout of EEPROM
basic unit
Basic unit EEPROM defective.
Contact INFICON customer service.
W104
An EEPROM parameter
is initializing
• A new parameter was
introduced by a software
update.
• If the message occurs
constantly during running up,
the EEPROM on the basic unit
is defective.
• Confirm the warning.
• Check whether the factory
settings correspond with the new
parameter of your application.
• Contact INFICON customer
service.
W106
EEPROM parameter
initializing
• A software update introduced
new parameters.
• The motherboard was
replaced.
• If the message occurs
constantly during running up,
the EEPROM on the basic unit
is defective.
• Confirm the warning.
• Check whether the settings
correspond with your application.
• Please contact INFICON
customer service
E107
Internal IIC
communication error
Internal defect
Please contact INFICON customer
service
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......