INFICON
Operation | 6
HLD6000-Operating-instructions-kinb43en1-05-(2107)
41 / 90
4
> Setup > Interfaces > Device selection
5
Select "Bus" as the module at M12 connector.
6
Confirm with
.
6.2.9.2 Setting a bus module address
You can set the setpoint for the field bus address. With PROFIBUS® this is the node
address. DeviceNet is the MAC ID.
This setting is irrelevant for Profinet and EtherNet/IP.
1
> Setup > Interfaces > Bus module > Adress
2
Use the keyboard shown for your entries.
3
Confirm with
.
ð
The value set is first carried over when restarting the HLD6000. To do this,
switch the power supply off and back on.
6.2.10 Setup scope of error messages
You can set up the scope within which error messages are shown on the display
during operation of the device. This can be set up separately for the supervisor and
the operator.
1
> Setup > Advanced setup > Error information
2
Under "Error information operator" or "Error information supervisor" choose
between
–
No.
–
No. and text
–
No. , text and info
3
Confirm with
.
6.2.11 Access to the settings
Role concept
• If you are logged in as a "supervisor” then you can change all settings available in
the device, protect against changes and check test objects for leaks.
• If you are logged in as an "operator" then you can change settings as far as is
authorized and check test objects for leaks. The scope of the adjustment options
depends, on the one hand, on the factory settings and on the other hand whether
the "supervisor" has assigned or withdrawn permissions. When individual
parameters are grayed out then no changes can be made.
• For a list of the preset parameter access controls, see “Factory settings [
The active role can be seen on the menu bar on the touchscreen:
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......