3 - 35
IP
N 07
4-
39
7-
P1
G
HAPSITE Smart Operating Manual
CAUTION
The concentrator feature is designed to enhance
HAPSITE sensitivity when analyzing low concentration
samples. The accumulated sample VOCs are transferred
from the concentrator adsorbent bed to the HAPSITE as
a concentrated sample. Care should be taken to avoid
overloading the HAPSITE and causing excessive
carryover of VOCs in subsequent runs.
3.6.1 Quick Reference SOP — Tenax Cleanout
1
From the main menu on the Front Panel highlight
Run Method
and press
SEL
.
2
Highlight
GCMS
and press
SEL
.
3
Highlight
Concentrator
and press
SEL
.
4
Highlight
3 min Tenax Conc Cleanout
and press
SEL
.
5
Wait for heaters to reach set temperatures.
6
When prompted press
RUN
from the Front Panel or Probe.
NOTE:
If the
TIC_Max
count is less than 500,000 at the end of this run, proceed
to the
Air 15 min Tenax Conc
method. If the
TIC_Max
count is greater
than 500,000, rerun the
3 min Tenax Conc Cleanout
method. Repeat this
until the
TIC_Max
is less than 500,000.
3.6.2 Quick Reference SOP — Tenax Concentrator Method
1
The 3 min Tenax Conc Cleanout method must be run prior this method. Refer
to
Section 3.6.1, Quick Reference SOP — Tenax Cleanout, on page 3-35
.
2
From the main menu on the Front Panel highlight
Run Method
and press
SEL
.
3
Highlight
GCMS
and press
SEL
.
4
Highlight
Concentrator
and press
SEL
.
5
Highlight
Air 15 Min Tenax Conc
and press
SEL
.
6
Wait for heaters to gain control of set temperatures.
7
When prompted press
RUN
from the Front Panel or Probe.
8
Hold the probe over the sample for both the one minute
LinePurge
and the one
minute
Concfill
event.
9
Wait for run to complete to view results.
Summary of Contents for HAPSITE Smart
Page 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Page 2: ......
Page 12: ......
Page 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Page 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Page 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 548: ......