19
Next, inject the calibration gas at a flow rate of 30 l/h (See Annex 2).
Wait for the signal to stabilize and adjust the signal with the sensitivity potentiometer (Ref. 2, Fig.
19).
The output signal should be:
EXAMPLE:
Sensor measure (% of full
range)
Output signal (mV)
0
880
50
2640
100
4400
Stop injecting the calibration gas and verify that the reading returns to zero (880 mV). If it does not,
repeat the procedure.
CALIBRATION COMPLETE
Flip the switch (Ref. 1, Fig. 17) into the “MES” position.
1.6. CTX300 CO
2
calibration
Warning: the sensor should be turned on for 15 minutes before adjustments are made.
The following text describes the steps necessary to adjust the transmitter (first calibration)
Fig. 20
(P 5)
Ref. 1
Ref. 2
V
Fig. 19
Ref. 3
U = 880 mV + 3520 mV
Sensor measurement range
S3
S2
S1
1
2
3
ISC Devices
Summary of Contents for OLDHAM 300 Series
Page 20: ...21 CALIBRATION CURVES CO2 IR Transmitter Output Signal...
Page 21: ...22 CO2 DETECTOR 0 10000 ppm 1 measurement range DISPLAY Display Concentration in ppm CO2...
Page 22: ...23 CO2 DETECTOR 0 5 CO2 measurement range DISPLAY Display Concentration in CO2...
Page 23: ...24 CO2 DETECTOR 0 10 CO2 measurement range DISPLAY Display Concentration in CO2...
Page 24: ...25 CO2 DETECTOR 0 50 CO2 measurement range DISPLAY Display Concentration in CO2...
Page 34: ...35 VI CERTIFICATIONS...
Page 35: ...36...
Page 36: ...37...
Page 41: ...42 ANNEX 3...
Page 42: ...43...
Page 43: ...44...
Page 44: ...45...
Page 45: ...46...
Page 46: ...47...
Page 47: ...48...
Page 48: ...49...
Page 49: ...50...