1.5. CSC 300 (semiconductor) calibration
Flip the switch (Ref. 1, Fig. 17) into the “CAL” position.
Ensure that the sensor is in clean air, otherwise inject synthetic air into it using the calibration kit
and referring to the recommendations below:
Important:
to correctly calibrate a sensor equipped with a semi-conductor cell, use of a
humidifier kit is MANDATORY (ref:
6335919) – Fig. 18
USING THE HUMIDIFIER KIT
Lift the lid (Fig. 18 ref. A) and, using a washbottle, moisten the filter, without saturating it, with
distilled water.
Replace the lid and check that all parts are properly assembled and that the assembly is fully
airtight.
Adjust the flow rate to 60 l/h and wait 10 minutes until the humidifier is fully purged.
Apply the gas introduction pipe to the nose of the sensor and wait at least five minutes for the
measurement to stabilize.
Note
: the sensor must be powered for at least two hours before any adjustment can be made.
Connect a voltmeter as indicated (Fig. 19 – Ref. 3) and adjust, using potentiometer P5 (Fig. 19, Ref.
1). The output signal must be equal to
880 mV
.
Rep 1
Fig. 17
Circuit du capteur CTX/CSC 300
MES
CAL
Vers capteur
Entrée gaz
Fig. 18
Gas
To detector
Summary of Contents for OLDHAM 300 Series
Page 20: ...21 CALIBRATION CURVES CO2 IR Transmitter Output Signal...
Page 21: ...22 CO2 DETECTOR 0 10000 ppm 1 measurement range DISPLAY Display Concentration in ppm CO2...
Page 22: ...23 CO2 DETECTOR 0 5 CO2 measurement range DISPLAY Display Concentration in CO2...
Page 23: ...24 CO2 DETECTOR 0 10 CO2 measurement range DISPLAY Display Concentration in CO2...
Page 24: ...25 CO2 DETECTOR 0 50 CO2 measurement range DISPLAY Display Concentration in CO2...
Page 34: ...35 VI CERTIFICATIONS...
Page 35: ...36...
Page 36: ...37...
Page 41: ...42 ANNEX 3...
Page 42: ...43...
Page 43: ...44...
Page 44: ...45...
Page 45: ...46...
Page 46: ...47...
Page 47: ...48...
Page 48: ...49...
Page 49: ...50...