Operation Guide
32
INDUSTRIAL SCIENTIFIC
Rev 2.0 (p/n: 17140724)
Zero Enable Screen
The Zero Enable screen displays the zero
icon, the up and down arrow icons, the lock
icon, the current zero enable value (0=do
not allow zeroing from the main monitoring
mode and 1=allow zeroing from the main
monitoring mode).
Use the Enter button to toggle between these values. A “0” value is
disabled. A “1” value is enabled. If zero in field is not enabled, calibration
in field is automatically also not enabled.
After the desired setting is selected, press the On/Off/Mode button to
display the Cal Enable screen.
Cal Enable Screen
The Cal Enable screen displays the gas
bottle icon, the up and down arrow icons,
the lock icon, the current cal enable value
(0=do not allow calibration from the main
monitoring mode and 1=allow calibration
from the main monitoring mode).
Use the Enter button to toggle between these values. A “0” value is
disabled. A “1” value is enabled.
After the desired setting is selected, press the On/Off/Mode button to
display the Cal Due Alarm Option screen.
Cal Due Alarm Option Screen
The Cal Due Alarm Option screen displays
the cal gas bottle icon, the calendar icon,
the alarm icon, the up and down arrow
icons, and the current setting for this
parameter (0=disabled, 1=enabled: alarm
occurs when calibration is due). Use the
Enter button to toggle between these
values.
After the desired setting is selected, press the On/Off/Mode button to
display the Cal Due Setpoint screen.