High Alarm Set Point
The High Alarm Set Point screen displays
the four high alarm set points, the alarm
icon, and the up arrow (high limit) icon.
If no changes are needed, press the
On/Off/Mode button to move to the next
screen (the TWA Alarm Set Point screen).
If changes are desired, press the Enter button. The first high alarm value
flashes. To adjust this value, use the Enter button. Once the desired value
is met, press the On/Off/Mode button to select the next high alarm value.
Continue this process until all four high alarm set points have been set.
Once all four values are set, press the On/Off/Mode button to submit the
values. Pressing the On/Off/Mode button again displays the TWA Alarm
Set Point screen.
TWA Alarm Set Point (Optional)
The TWA Alarm Set Point screen displays
the TWA alarm set points (for MX4s
configured with toxic gas sensors), the TWA
icon, and the up and down arrow icons.
If no changes are needed, press the
On/Off/Mode button to move to the next
screen (the TWA Interval screen).
If changes are desired, press the Enter button. The first TWA alarm value
flashes. To adjust this value, use the Enter button. Once the desired value
is met, press the On/Off/Mode button to select the next TWA alarm value.
Once the values are set, press the On/Off/Mode button to submit the
values. Pressing the On/Off/Mode button again displays the TWA Interval
screen.