
28
29
Problem
Cause
Corrective measures
The device is not
turning on.
This can occur when
the power cable is
not fully inserted
into the power outlet.
•
Plug the power cable into a grounded 3-prong outlet.
•
When using a power strip, make sure the switch is turned on.
•
Make sure the power cable (AC 250V 10A) provided by
InBody is in use.
Power strip
IV.
FAQ
If a problem persists, please contact InBody.
A. Error Code
The error messages can be displayed on the screen if they occur during the test.
Please provide the error code to Product Support.
B. Regarding the Equipment
If problems cannot be resolved with the information below, please contact InBody.
3-prong outlet
Type B
<Plug Type B>
<Plug Type F>
<Plug Type B>
<Plug Type F>
<Plug Type B>
<Plug Type F>
Power strip
3-prong outlet
Type F
Summary of Contents for BPBIO320
Page 4: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ...inbody com 2022 InBody Co Ltd All rights reserved IM ENG 82 L 220629 1639 ...