II. InBody Test

  A. Precautionary Steps

                                

• 

Individuals with medical implant devices such as pacemakers, or essential support devices such as patient 

  monitoring systems, must not use this equipment. Safe, low-level currents will flow through the body during 
  the test, which may cause malfunctioning of the device or endanger lives.

• 

Children and people with limited mobility should be supervised or assisted when attempting to test on the InBody.

• 

After an individual with any kind of contagious disease or infection tests on the InBody, use an InBody Tissue

  to clean the equipment.

Warning

• 

Stand upright for about 5 minutes before testing. Taking the test immediately after lying in bed or sitting for

   a long period of time might result in a slight change in the test results. This is because body water tends to
   move to the lower body as soon as the person stands or gets up.

• 

Do not eat before testing. In cases where the examinee has already eaten, the test should be put off for at least

   two hours after the meal. This is because food mass is included in the examinee’s weight and thus, may result
   in measurement errors.

• 

Use the bathroom before testing. Waste is not included in the body’s compositional elements, but the volume 

  of urine and excrement is included in the weight measurement affecting accuracy of the test results.

• 

Do not exercise before testing. Strenuous exercise or sharp movements can cause temporary changes in body

   composition. Even light exercise can change your body composition temporarily.

• 

Take the test in the morning, if possible. Body water tends to gravitate towards the lower body throughout

  the day, affecting accuracy of the test results.

• 

Thoroughly wipe the palms and soles with the InBody Tissue before testing. Testing may be difficult if the 

  examinee’s palms and soles are too dry or if the examinee has too many calluses. 

• 

Avoid contact with the examinee during testing. Contact may lead to interference affecting test results.

Caution

370S_manual_Eng_C_181224.indd   16

2018. 12. 24.   오후 7:29

Summary of Contents for 370S

Page 1: ...User s Manual 370S_manual_Eng_C_181224 indd 1 2018 12 24 오후 7 29 ...

Page 2: ... product can also be used for medical purposes as the InBody370S measures body composition Representative Sponsor Information InBody Europe B V Hessenbergweg 75A 1101CX Amsterdam Zuidoost The Netherlands TEL 31 20 2386080 Website https nl inbody com E mail info eu inbody com Australian Sponsor AUSTRALIA Emergo AUSTRALIA Level 20 Tower II Darling Park 201 Sussex Street Sydney NSW 2000 AUSTRALIA TEL...

Page 3: ...uctions C Test Posture III Transportation and Storage A Cautions during Transportation B Repacking Instructions C Transportation and Storage Environment IV Frequently Asked Questions FAQ A Regarding the InBody B Regarding the InBody Test V Others A Exterior and Functions B Safety Information C Classification D Specifications E EMC declaration 4 5 5 8 12 15 16 17 18 19 19 19 20 22 23 26 28 28 31 Us...

Page 4: ... the following components are present Please inspect each component of the InBody370S for defects prior to installation InBody370S Adapter DC 12V 3 4A 1 EA Power cord 1 EA User s Manual 1 EA Keypad Hand electrodes Upper part Footplate Foot electrodes 370S User sManual 제품보증서 370S_manual_Eng_C_181224 indd 4 2018 12 24 오후 7 29 ...

Page 5: ...please contact InBody for assistance Do not transport the equipment by holding the screen portion or the joints of the hand electrodes Keep the packing materials provided for repacking the equipment in the future Other wastes should be disposed of according to relevant laws and regulations Caution Using the InBody370S on carpet may cause static electricity which could damage the equipment If insta...

Page 6: ...ator Leveling screws under the footplate Leveling screws under the footplate Power input port Rear panel 3 socket outlet Grounded outlet Avoid injuring your hands when rotating the leveling screws under the footplate Caution Do not place the InBody370S in a location making it difficult to disconnect the power cord Do not plug in or pull out the power cord with wet hands There is a risk of an elect...

Page 7: ...rge AC motor equipment treadmill vibration plate refrigerator air conditioner compressor etc high frequency thermal therapy equipments or heating appliances Do not share the power source of the InBody370S with other electrical devices This may affect the test results When connecting the InBody370S with other test equipment turn on the other equipment first When turning off other equipment turn off...

Page 8: ...ate Please do not stand on the footplate or place objects on the footplate 2 Press the Administrator Menu button on the screen which appears when no one is on the footplate 3 Input the password default password 0000 to access the Administrator Menu 4 The Administrator Menu will give you access to Setup and Troubleshooting Administrator Menu Setup Troubleshooting 370S_manual_Eng_C_181224 indd 8 201...

Page 9: ...tion Combine Back up InBody Test results to a USB Thumb Drive or restore test results using a backup file on a USB Thumb Drive or add test results using a backup file on a USB Thumb Drive to the InBody 09 Printer Setup Connect the printer to the InBody A connected printer will allow for printing results sheets after testing 10 Results Sheet Types Select which results sheets to utilize with the InB...

Page 10: ...t measured weight by a fixed value on the InBody Example Workout clothes at the gym are approximately 0 2kg most examinees are assumed to be wearing workout clothes so the examiner may adjust the set value to 0 2kg 20 Normal Range Set the normal range for BMI Percent Body Fat and Waist Hip Ratio The ideal value of BMI may also be set 21 N A 22 Standard Child Growth Curve Set the type of standard c...

Page 11: ... 02 Results Sheet does not print View the troubleshooting checklist when the Results Sheet does not print by the printer connected to the InBody 03 Weight is not being measured View the troubleshooting checklist when weight is not being measured after stepping on to the InBody footplate 04 Weight measurement seems to be inaccurate View the troubleshooting checklist when the weight measurement seem...

Page 12: ... You can edit your printing settings under Settings of the Administrator Menu from 10 Results Sheet Types through 15 Printing Alignment 2 Thermal Printer Connect thermal printer to the InBody370S to print Thermal Results Sheet Always connect a thermal printer from InBody 1 Turn off the InBody370S For best result the InBody370S should be turned off when connecting to thermal printer 2 Plug the seri...

Page 13: ...d pressure monitor will be sent directly to the InBody370S Always connect a blood pressure monitor from InBody If you select to print blood pressure measurements under Settings of the Administrator Menu 13 Outputs Interpretations for Results Sheet the blood pressure measurements will be printed on the InBody Results Sheet 1 First turn off the InBody370S and then the blood pressure monitor You may ...

Page 14: ...tion issues in connecting the barcode reader to the InBody370S if the InBody is turned on 2 Plug the USB cable of the barcode reader into the USB HOST port on the rear panel of the InBody370S 3 Turn on the InBody370S If the barcode reader is connected to the InBody370S the barcode reader icon will appear on the top left corner of the screen when no one is on the footplate USB HOST port 370S_manual...

Page 15: ...he equipment directly Keep food and drinks away from the equipment Substances getting inside the equipment can cause critical damage to the electronic components Use a lint free cloth to gently wipe the external surface of the equipment about once every week Be careful not to scratch the LCD screen InBody370S does not need regular maintenance If some problems occur while operating the device get i...

Page 16: ... where the examinee has already eaten the test should be put off for at least two hours after the meal This is because food mass is included in the examinee s weight and thus may result in measurement errors Use the bathroom before testing Waste is not included in the body s compositional elements but the volume of urine and excrement is included in the weight measurement affecting accuracy of the...

Page 17: ...examinee takes the InBody Test following the instructions that are displayed on screen 2 Weight measurement begins 3 Input personal information Input height only if using Self Mode In Professional Mode enter the ID 4 Maintain proper posture to take the test Refer to C Test Posture for the proper posture 5 The InBody Test begins InBody Information is shown if using Self Mode Professional Mode Self ...

Page 18: ...rical contact Professional Mode Self Mode Place the heels on the rear sole electrodes Keep your arms straight The thighs should not touch Hold the hand electrode so that the 4 fingers wrap the surface of the bottom hand electrode and place the thumb on the oval electrode Your arms must not touch the sides of your body Step on the footplate barefoot 370S_manual_Eng_C_181224 indd 18 2018 12 24 오후 7 ...

Page 19: ...S with the polyethylene foam cover After covering the InBody lower the upper part 3 Put the InBody into the packing box Place the packing pad over the equipment and tape up the packing box C Transportation and Storage Environment 1 The InBody370S should be transported or stored under the following conditions Temperature range 10 70 C 14 158 F Relative humidity 10 80 RH No Condensation Atmospheric ...

Page 20: ... resolved through the Troubleshooting please refer to the possible solutions below Question Answer My InBody370S does not turn on Insert the power plug completely into a grounded 3 socket outlet When using a power surge protector the equipment may not power on if the power switch on the power surge protector is turned off Check the power surge protector which the power plug is connected to The pro...

Page 21: ...he last touchscreen alignment and will automatically recall the previous touchscreen settings The InBody can also recognize if its touchscreen alignment is off screen and will automatically take the user to the calibration screen after restarting I would like to connect other equipment to the InBody370S Please refer to E Connecting Printer Thermal Printer Stadiometer Blood Pressure Monitor and Bar...

Page 22: ...ls with metallic implants may have skewed test results due to the conductivity of the metal affecting the results As the weight of clothes and other wear affects the results of the body composition analysis it is strongly recommended to take off any heavy clothing or metallic wear Except for the weight jewelry does not effect the body composition analysis as the contact point with the InBody370S a...

Page 23: ...placed on the oval electrode Hand electrode joint and handle Supports the hand electrode and encloses the wiring for the electrode Body Connects the upper part of the equipment to the lower part LCD screen Shows each stage of the test instructions test results etc You can touch the screen to input the data required for the test configure settings or view test results Number keypad Used for inputti...

Page 24: ...ctrode by stepping with the heel of their foot Footplate This is connected to the scale which measures the examinee s weight Level Indicator Indicates the current horizontal level of the InBody370S Joint Joins the upper part and lower part of the equipment together Leveling screws Used for adjusting the horizontal level of the equipment 370S_manual_Eng_C_181224 indd 24 2018 12 24 오후 7 29 ...

Page 25: ...essure monitor serial port Female RS 232C Used for connecting the InBody370S to a blood pressure monitor Only compatible with a InBody blood pressure monitor LAN port 10T Base Used for connecting the InBody370S to Lookin Body installed on a computer The InBody370S can be connected to Lookin Body installed on a computer using one of the ports or USB SLAVE port Used for connecting the InBody370S to ...

Page 26: ...10T Base USB port 4 4 4 Do not touch signal input signal output or other connectors and the patient simultaneously External equipment intended for connection to signal input signal output or other connectors shall comply with relevant IEC Standard e g IEC60950 for IT equipment and IEC60601 1 series for medical electrical equipment In addition all such combination system shall comply with the stand...

Page 27: ...bout recycling this product please refer to local governing ordinances and recycling plans Follow instructions for use Suivez les instructions d utilisation WARNING Electric shock hazard do not dismantle Dismantling will void the warranty AVERTISSEMENT Risque de choc électrique ne pas démonter Le démontage annulera la garantie DANGER Do not use this equipment with electrical medical device such as...

Page 28: ...dy Composition History Weight Skeletal Muscle Mass Percent Body Fat InBody Score Body Type Based on BMI Percent Body Fat Athletic Shape Slightly Obese Obesity Muscular Shape Average Slightly Obese Slim Muscular Slim Sarcopenic Obesity Thin Slightly Thin Weight Control Target Weight Weight Control Fat Control Muscle Control Nutrition Evaluation Protein Minerals Fat Mass Obesity Evaluation BMI Perce...

Page 29: ...y saved settings changes Database Test results can be saved if the member ID is utilized The InBody can save up to 100 000 results Test Mode Self Mode Professional Mode Administrator Menu Setup Configure settings and manage data Troubleshooting Additional information to help use the InBody USB Thumb Drive Copy backup or restore the InBody test data data can be viewed on Excel or Lookin Body data m...

Page 30: ...onment 10 40 C 30 75 RH 70 106 kPa Storage Environment 10 70 C 10 80 RH 50 106 kPa No Condensation Testing Weight Range 10 250kg 22 0 551 2lbs Testing Age Range 3 99 years Height Range 95 220cm 3ft 1 40in 7ft 2 61in 370S_manual_Eng_C_181224 indd 30 2018 12 24 오후 7 29 ...

Page 31: ...ontact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge ...

Page 32: ...ermined by an electromagnetic site survey a should be less than the compliance level in each frequency range ᵇ Interference may occur in the vicinity of equipment marked with the following symbol NOTE1 At 80 MHz and 800 MHz the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures o...

Page 33: ...370S_manual_Eng_C_181224 indd 33 2018 12 24 오후 7 29 ...

Page 34: ...370S_manual_Eng_C_181224 indd 34 2018 12 24 오후 7 29 ...

Page 35: ...370S_manual_Eng_C_181224 indd 35 2018 12 24 오후 7 29 ...

Page 36: ...www inbody com 2018 InBody Co Ltd All rights reserved BM ENG I5 C 181224 370S_manual_Eng_C_181224 indd 36 2018 12 24 오후 7 29 ...

Reviews: