or
No display
Message
Correction
The temperature
measured was higher
than the specified
measuring range.
Please
measure again.
The temperature
measured was lower
than the specified
measuring range.
Please
measure again.
The am bi ent (room )
tem per at ur e was out si de
the oper at ing range.
Keep
it at roo m tem perat ure
(app rox. 16
°C
~40
°C
) for
30 minu tes befor e usi ng
.
Replace the two
"AAA" batteries
Error Message
This thermometer is guaranteed for
one year from the date of purchase
against manufacturer's defect under
normal, household use . Following the
instruction closely will ensure years of
dependable operation.
If the thermometer does not function
properly, first check the batteries (see
battery replacement instruction) and
replace them if necessary.
Limited Warranty
14
Failed self-diagnostic test.
Reinstall batteries and try
again.
12
13
Please do not
modify or disassemble
the unit.
Please use a cotton
swab with alcohol to
gently clean the surface
of the lens. Before next
measurement, wait at
least 40 minutes for the
unit to dry.
Cleaning and Care
Please avoid thinner
and benzene or other
harsh cleaners.
Please remo ve the
batteries if the unit will
not be used for a long
time .
Please keep the unit in a
suitable place avoiding
high temperature, direct
sunlight, high moisture,
and dust. The unit is not
water proof.
Please do not drop the
unit. It is not shock-proof.
Warning: following
mechanical shock, the
unit should not be used
before re-calibration.
Z
Z
Z
Power Source
Two "AAA" (1.5Vx2) batteries
Measuring Range
Ear Mode: 0.0
~100.0
(32
~212
)
Forehead Mode:
32.2
~43.3
(90
~109.9
)
Ambient Mode : 10.0~40.0
,(50~104
)
Resolution
0.1
(
)
Accuracy
0.2
(0.4
) for blackbody temperature
range 35.5-42.0
(95.9-107.6
) ; Other
ranges
±
2%
Memory
10 sets
Low Battery
Replace the batteries if the low battery
indicator appears.
Auto Power Off
Auto shut off after one minute of non-use
Battery Life
Approximately 1000 measurements.
Operating Environment
16.0~40.0
(60.8~104.0
, < 95% RH
Storage Environment
-25.0~55.0
, (-13~131
),< 95% RH
103x51x40mm
Weight
°C
°C
°C
°F
°C
°C
°F
°F
°C
°F
°C °F
°C
°F
°C
°F
°C
°F
)
°C
°F
Approximately 62g (including batteries)
Dimensions
Guarantee of Quality
Certification ISO 13485
Conformity EN 12470-5 & IEC 60601-1
ASTM E 1965-98
Specifications
This infrared thermometer meets
requirements established in ASTM standard
(E 1965-98). Full responsibility for the
conformance of this product to the standard
is assumed by K-jump .