7. Click Write frequency and options.
8. Close the window.
Set Wake Up Events
The Tube can be programmed with up to 8 different wake up periods. A wake up period
is when the Tube sends data, sends a status text message, or waits for an external
connection.
Wake Up events are not necessary for most use cases.
1. wake up events.
2. Click the Wake up events tab.
3. Under "Wake up no. 1," select the amount of time for the Tube to wake up.
4. Select the day of the week for the wake up event.
5. Select the starting time. The Tube cannot be programmed for 12:00:00 AM.
6. Select the type of task for this wake up event. If "Wait for phone connections" is
selected, additional options become available.
7. Scroll down to program additional wake up events.
8. When all events are programmed, click Write frequency and options.
9. Close the window.
Restart Device and Test Communications
1. Click Functional mode and test.
2. Click the Probes, real time reading tab.
3. Press the Immediate reading probes button. Ensure all values are appropriate for
the current conditions of the instrument. If the readings are all zeros there is a
connection issue with your instrument. See "Changing Instrument Communication
Settings" on page 8.
4. Click the Test tab.
5. Click the FTP test button to have a sample data file sent to the FTP site you
specified in the previous setup.
6. Click the Test data email button to have a sample data file sent to the email
addresses you specified in the previous setup.