NOVUS AUTOMATION 

1/6 

 

 

LogBox-RHT-LCD 

TEMPERATURE AND HUMIDITY LOGGER – INSTRUCTIONS MANUAL V2.0x A 

1

 

PRESENTATION 

LogBox-RHT-LCD

  is an electronic temperature and relative 

humidity logger. Sensors measure these quantities and the values 

obtained are visualized on the LCD display and stored in electronic 

memory. The stored data is later sent to a computer, which allows it 

to be viewed and analyzed in the form of tables or graphs. It is 

possible to configure the humidity output to display  the measured 

Relative Humidity

 value or the 

Dew Point

 temperature value. 

The 

NXperience

 software is the tool used to set up the device and 

view the data downloaded  by the logger. Parameters such as 

acquisition start and end times and intervals between acquisitions, for 

example, can be easily configured through the 

NXperience

 software. 

The data acquired by the logger can be exported for analysis in other 

programs, such as spreadsheet programas. 

1.1

 

IDENTIFICATION 

The identification label is attached to the logger housing. Verify that 

the characteristics described on this label are in accordance with 

what was requested. The 

RHT-LCD

 model measures temperature 

and relative humidity. 
The following elements are shown in the logger front: 

 

Figure 1 –

 

Logger Front 

START  /  STOP  button

:  Button that can be configured to start or 

stop acquisitions. When pressed, it also used to navigate or change 

screens. 

Status  Indicator  (LOG)

:  This  indicator  will be active when the 

logger is performing acquisitions and will be deactivated when the 

acquisitions are finished or while they are not started. 

Alarm Indicator  (AL1  and/or  AL2)

:  This  indicator  informs you of 

alarm situations that have occurred and will be active whenever an 

alarm situation occurs. It will remain in this condition until a new 

configuration is applied to the logger. 

IR Communication Window  (COM)

:  This  indicator  is activated 

when a communication attempt is made or when communication 

between the logger and the computer is established. 
 

2

 

SPECIFICATIONS 

Operating 

temperature 

From -40 °C to 70 °C 

Sensor 

measurement 

range 

Temperature: -40.0 °C to 80.0 °C. 
Relative Humidity (RH): 0.0 to 100.0 %, 

without condensation. 
Dew Point: -40.0 °C and 100.0 °C 

Accuracy 

See 

Figure 2

Note

:  In the 

Offset

  parameter of the 

NXperience

 software, you can erase the 

measurement error found. 

Measurement 

resolutions 

Temperature: 0.1 °C 
14 bits (16383 levels) 
Relative Humidity (RH): 0.1 % 
12 bits (4095 levels) 

Response time 

Temperature:  Up  to 30  seconds  with 

fairly still air. 

Humidity: Up to 8 seconds with fairly still 

air (20 to 80 % RH). 

Memory capacity 

32.000 (32 k) logs: 

Half

 for each channel 

or 

total

 when one of the channels is not 

enabled. 

Measurement 

interval 

1 second min. 18 hours max. 

Power supply 

3.6 Vdc lithium battery (1/2 AA), built-in 

Estimated 

autonomy 

Higher than 200 days  with weekly data 

reading.  
Frequent logged data readings may 

shorten battery life and the interval 

between acquisitions is too short. 

Housing 

Polycarbonate 

Protection 

Suitable for products with protection level 

IP40

Dimensions 

60 x 70 x 35 mm 

Logger-PC data 

transfer time 

According to the number of logs. 40 

seconds for 16.000 logs. 

PC interface 

Interface IR/USB or IR/Serial. 

NXperience 

software 

Configurator software for Windows 10, 8, 

7 and XP. 
Menus in Portuguese, English, French or 

Spanish. 

Summary of Contents for LogBox-RHT-LCD

Page 1: ... finished or while they are not started Alarm Indicator AL1 and or AL2 This indicator informs you of alarm situations that have occurred and will be active whenever an alarm situation occurs It will remain in this condition until a new configuration is applied to the logger IR Communication Window COM This indicator is activated when a communication attempt is made or when communication between th...

Page 2: ...sures For skipping idle mode press button for activating communication temporarily After an acquisition configuration has been sent it will definitely be out of idle mode Battery Saving Mode MAIN SCREEN Indication screen for values measured by the register 1st Screen Temperature and Humidity 2nd Screen Temperature and dew point INDICATOR SCREEN MIN MAX CH1 It indicates the minimum and maximum valu...

Page 3: ...l is equal to 10 minutes and disp screen is in lo6 mode in every 1 minute a measure will be made to make the 10 necessary measures for an acquisition In this case consumption will be same as Ex 2 Example 4 Starting and ending of acquisitions It has to observe the best start and end acquisitions mode to save battery due to necessity of the measures If the process in which the measures are performed...

Page 4: ...perature supported by the device Upper Alarm Allows you to enable and configure a higher alarm setpoint option available only if the Start Mode is set to Immediate Start or Date Time o Maximum Displays the maximum temperature supported by the device Lower Setpoint When the measured temperature is lower than the value set in this parameter the device will start the logs option available only if the...

Page 5: ...t be interrupted following the device configuration 7 4 1 DOWNLOADING DATA Data download is accomplished by clicking the button Download Logs in the Download screen from NXperience During data transference a status bar indicates remaining data to be transferred Data downloading time is proportional to the number of readings logged 7 4 2 VISUALIZING DATA At the end of values transfer the data will ...

Page 6: ...act with the electronic circuitry due to the risk of damage from static electricity Carefully check the battery polarity When closing the housing the cover must be properly tightened back in its place such as to assure the sealing protection Used batteries should not be recharged dismantled or incinerated After use batteries must be disposed according to local legal rules or returned to supplier N...

Reviews: