12
www.in-situ.com
3. Click the Immediate reading probes button. Note the units used for level readings.
4. Close the window.
5. Click Probe and set-point configuration.
6. Click the Level reference setting tab.
7. Select the instrument to change the level reference from the drop down menu.
The instrument must be connected by cable to the Tube to change
the level reference.
8. Select the appropriate level mode in the Parameter Id drop down menu. Refer to
your instrument manual for more information on level modes.
9. Click Save Id parameter.
10. Enter your level reference value.
11. Click Set the Level reference button.
12. Close the window.
Set the Site Name and Clock
1. Click General information, parameters and functional options.
2. Click the Information tab.
3. Enter a site name for your deployment area and click Set the Site Name.
4. Click Set clock to synchronize the Tube clock with your computer clock.
5. Close the window.
Configure Options
1. Click General information, parameters and functional options.
2. Click the Options tab.
3. "Barometric compensation for level readings" is only changeable if you have a Tube
with an internal barometric sensor. Enter a manual value for the density of water if
needed.
4. Select "Power supply to the probes" if you are deploying an Aqua TROLL 400, RDO
PRO-X probe, or Rugged TROLL 200.
5. Select whether to send data to an email address or to an FTP site.
Do not alter the settings in the data file settings.