background image

OPERATING INSTRUCTIONS

1. Zero the meter

Prior to use, the meter must be zeroed. To do this, ensure the 
meter is in a non-pressurized area, and press and hold the 
‘ZERO’ button for approximately four seconds. The display will 
show ‘888’ to indicate that the zero function has occurred.

2. Measure Pressure

Connect the flexible tube between the meter and the 
pressure source.  

The EM201’s positive input is to the right when looking at the 
front of the instrument. Applying a positive pressure to this 
port will display a positive reading. The input to the left is the 
negative port. (Applying a positive pressure here will display a 
positive reading). Use either input to measure a single source 
by venting the unused input to atmosphere. The instrument will 
continually provide a differential reading of the two inputs.

4

Summary of Contents for EM 151

Page 1: ...ental Equipment Inc 3634 Central Ave St Petersburg Fl 33711 Phone 727 328 2818 or 1 800 RING IMR Fax 727 328 2826 E mail info imrusa com Web www imrusa com INSTRUCTION MANUAL ELECTRONIC MANOMETER EM 151 EM 201 ...

Page 2: ...Chart 7 Warranty 8 INTRODUCTION The EM151 and EM201 provide a simple time saving alternative to U tube pressure measurements Use the UEi EM151 EM201 to measure gas supply and draft pressures in residential or light commercial properties The EM201 differential mode provides a convenient method to measure pressure drops ...

Page 3: ...y and ventilate the occupied space Identify and repair the source of the leak before continuing with the test In the interest of safety this instrument should only be used by trained competent professionals who know the hazards and consider the risks of working on and with industrial equipment These meters do not measure water pressures under these conditions they will fail Do not exceed 10 PSI in...

Page 4: ...on Measures in inches water gauge Auto power off after 20 minutes 3 year limited warranty EM201 Dual input differential measurement to 60 wg Triple Scale inWG mBar and PSI Ruggedized including rubber boot with integral magnet Measures in inches water gauge and mBar Auto power off after 20 minutes Min Max value capture Data Hold Back lit display 3 year limited warranty ...

Page 5: ...ct the flexible tube between the meter and the pressure source The EM201 s positive input is to the right when looking at the front of the instrument Applying a positive pressure to this port will display a positive reading The input to the left is the negative port Applying a positive pressure here will display a positive reading Use either input to measure a single source by venting the unused i...

Page 6: ...he meter off then back on Captured values can be displayed in either inches water gauge or mBar 5 HOLD EM201 only The EM201 has a display hold feature Press this during reading to freeze the display Press again to return to actual measurement 6 BACKLIGHT EM201 only The EM201 has a backlit display Press to turn on or off the back light feature 7 Over range If the input exceeds the meter range OL wi...

Page 7: ... 10 10 wg Operating range 0 40 C 10 90 RH non condensing Battery life 50 hours min with alkaline battery EM201 Measurement range 60 to 60 WG Minimum pressure displayed 0 01 WG Accuracy 0 02 2 WG 1 rdg 2 WG Resolution 0 001 9 999 to 9 999 WG 0 01 10 10 wg Operating range 0 40 C 10 90 RH non condensing Battery life 40 hours min with alkaline battery ...

Page 8: ...7 Multiply By To Get In of H2O 1 868 mmHG In of H2O 0 074 in HG In of H2O 0 249 kPa In of H2O 249 1 Pascals In of H2O 2 491 mBar In of H2O 0 036 PSI In of H2O 25 40 mmH2O PRESSURE CONVERSIONS ...

Page 9: ...ty materials required for the function of the unit and other wear parts unless they have a direct bearing on work performed under the warranty 6 The terms and conditions for the acknowledgement of this warranty shall be the presentation of the fully completed warranty card which must contain the confirmation from the authorized IMR dealer on its delivery and if applicable the prescribed maintenanc...