Models: ALN-T02-2L3S-GA
**IMPORTANT**
GUIDELINES TO ASSURE SAFE AND PROPER OPERATION OF YOUR TABLE
INSTALLATION GUIDELINES:
- Never open the A/C Control Box.
- Only use supplied Power Cable.
- Do not install table on an uneven surface.
- Do not install table in dusty or humid locations.
- Never lean on or apply excessive weight to table while in operation.
- Prevent wires from becoming caught in the mechanisms by securing them with the supplied clips and ties. Keep motor
area clear from wires and any other objects at all times.
- Only use a dry or damp cloth to clean Control Box. Before cleaning UNPLUG from power source.
- Check that all connectors are securely plugged in; remote wire and motor(s) wire(s) and A/C Control Box.
RESET INSTRUCTIONS
1) Hold down “
” button for 5 seconds.
2) The readout will show ‘ASF.’
3) Release the button.
4) Press and hold the “
”button. The table will descend to its lowest level, stop, and rise until reaching default height - the
readout will now show numbers instead of ‘ASF.’ DONE!
****LOWEST LEVEL POSITION MUST BE REACHABLE. Path must be clear of obstructions when lift table is descending,
it cannot hit a stop before lowest LEVEL position otherwise reference is stored at the wrong height. ****
AFTER RESET, run full UP and DOWN cycles several times to ensure problem is solved.
OTHER ISSUES AND SOLUTIONS:
If ONE LEG is higher than the OTHER:
1) Unplug Power
2) Swap the left motor connector with the right motor connector on the A/C Control Box.
Repeat RESET procedure
.
If there is no change, the problem is with the
MOTOR LEG
.
If the change results in the same problem but with the other leg, the problem is with the
A/C CONTROL BOX
.
Do not hesitate to contact us if you run into any problems during the procedure, or if you need a replacement:
MOTOR LEG
or
A/C CONTROL BOX
.
Thanks!