7
Explorer
Pro
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
Go Back
Help
Enter
On/Off
Go Back
Help
Enter
On/Off
PROCEDURE
• Press
arrow button to select MAIN MENU.
• Press
Enter
button.
• Using the arrow buttons, select CALIBRATION.
• Press
Enter
button.
• Press
or
arrow button and select AUTOCAL.
• Clear the pan.
• Press
Enter
button, screen displays PLEASE WAIT, the
balance is automatically calibrated and then returns to a
weighing mode.
NOTE:
AutoCal
TM
calibration uses an internal mass in the balance for
calibration and is done automatically when selected.
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
INTERNAL CALIBRATION
PLEASE CLEAR THE PAN
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
INTERNAL CALIBRATION
PLEASE WAIT . . .
0.0
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
0.0
WEIGHT
0%
100%
CONTRAST
GRAMS
STABLE
MOVE HIGHLIGHT
CONTRAST
MAIN MENU
3.3.1
Auto Calibration (AutoCal
TM
)
On
Explorer
Pro
Balances equipped with the AutoCal
TM
feature, calibration can be accomplished using an internal calibration
mass. Auto calibration can be performed at any time providing the balance has warmed up to operating temperature.
Summary of Contents for BH350