Datalogging with Differential Analog Inputs
Using a datalogger with a common ground will produce considerable problems, due to earth loops.
The higher the length of the cable the higher the problems with earth loops. Earth loops can produce
considerable fluctuations at the measurement signal 0 to 1V !
IMKO recommends to use a datalogger with differential inputs as shown in the following picture. With
differential inputs the problems with earth loops are eliminated.