![IMKO SONO-VARIO LD User Manual Download Page 29](http://html1.mh-extra.com/html/imko/sono-vario-ld/sono-vario-ld_user-manual_2047643029.webp)
29/44
2.1.5. Test run in the respective Measurement Mode
In the menu "
Test
" and the window "
Test in Mode CA or CF
" the measured moisture values
“MoistAve”
(Average) of the SONO probe are displayed on the screen and can be parallel saved in a
file. In the menu "
Test
" and the window "
Test in Mode CS
" the measured single measurement values
“
Moist
” (5 values per second) of the SONO probe are displayed on the screen and parallel stored in a
file. In
„
Test in Mode A“
single measurement values (without average) are displayed on the screen
and can also be stored in a file.
Attention:
for a test run in mode CA, CF, CS or A it must be ensured that the SONO
probe was also set to this mode (Measure Mode CA, CF, CS, A). If this is not assured,
the probe returns zero values.
Following measurement values are displayed on the screen:
MoistAve
Moisture Value (Average)
MatTemp
Temperature
Conduct
Radar-based-Conductivity RbC
TDRAve
TDR-Level (for special applications)
DeltaCount
Number of single measurements which are used for the averaging.
tpAve
Radar time (average) which corresponds to the respective moisture value.
By clicking „
Save
“ the recorded data is saved in a text file in the following path:
\
SONO-CONFIG.exe-Pfad\MD\Dateiname
The name of the text file
SN+yyyymmddHHMMSS.sts
is assigned automatically with the
serial number of the probe (SN) and date and time.
The data in the text file can be evaluated with Windows-EXCEL.