imc CRONOS-PL/-SL/compact - Getting Started, Version 5 R 7 - 2018-06-28
© 2018 imc Meßsysteme GmbH
8
General
1.3 Instruction manual
This instruction manual provides important notes on using the device. The safe working is conditional on
compliance with all safety measures and instruction specified.
Additionally, all accident prevention and general safety regulations pertinent to the location at which the
device is used must be adhered to.
This instruction manual exclusively describes
the device (hardware)
, not how to operate the
imc
software
! The instructions for the imc measurement software are provided in their own manual. Read
carefully the manual before beginning any work!
1.4 Liability limitations
All specifications and notes in the operating instruction manual are subject to applicable standards and
regulations, and reflect the state of the art well as accumulated years of knowledge and experience.
The manufacturer declines any liability for damage arising from:
·
failure to comply with the instructions provided,
·
inappropriate use of the equipment,
·
additionally, the general terms and conditions of the company imc Mess-Systeme GmbH apply.
1.5 Guarantee
Each device is subjected to a 24-hour "burn-in" before leaving imc. This procedure is capable of
recognizing almost all cases of early failure. This does not, however, guarantee that a component will not
fail after longer operation. Therefore, all imc devices are guaranteed to function properly for two years.
The condition for this guarantee is that no alterations or modifications have been made to the device by
the customer.
Unauthorized intervention in the device renders the guarantee null and void.
1.6 Before starting
Condensation may form on the circuit boards when the device is moved from a cold environment to a
warm one. In these situations, always wait until the device warms up to room temperature and is
completely dry before turning it on. The acclimatization period should take about 2 hours. This is
especially recommended for devices without ET (extended environmental temperature range).
We recommend a warm-up phase of at least 30 min prior to measure.
Existing ventilation slits must be kept unimpeded to avoid heat buildup in the device interior.
The devices have been designed for use in clean and dry environments. It is not to be operated in 1)
exceedingly dusty and/ or wet environments, 2) in environments where danger of explosion exists nor 3)
in environments containing aggressive chemical agents.