4.4.2  Using single iQ-LED modules for low light 

With the iQ-LED Control SW version 3.2.0, you now have the option to turn off single iQ-LED modules in your 

device. This option gives you another possibility to decrease the intensity inside your device. Please note that 

when using single modules in edge positions of your LE7 (especially LE7-4x), the uniformity will decrease 

significantly. If using single modules, it is recommended to only use the center-positioned iQ-LED modules. How 

to deactivate/activate single modules is described in chapter 4.5 of the SW manual.

 

4.4.3  Low-light using the LE7-E 

In the LE7-E device, the top center iQ-LED is combined with a 1,8 ND filter. When using only this module with 

the described workaround from chapter 4.4.2, you will have the possibility to reach intensities to as low as 0,25 

lx. For example, when considering both the ND filter value of 1,8 and a correction factor regarding the optical 

structure, a generated illuminant D with 120 lx (using all modules, calibrated with the TE291 calibration chart) 

should correspond to an intensity of approximately 0,25 lx when using only the iQ-LED module with the ND filter.  

To get a rough idea of which intensities will appear when using only the ND module, you can use a Density Filter 

Compensation value of 

2,68

 in the iQ-LED SW. Please see chapter 3.1.3 in the software manual for details.

 

Please note that it is impossible to measure such low intensities with the internal spectrometer. Instead, we 

recommend using an external radio lux meter to verify the correct intensity. 

 

5  ADDITIONAL INFORMATION 

5.1  Maintenance 

The spectrometer requires a recalibration yearly, regardless of the operating hours. If a spectrometer calibration 

is necessary, please contact Image Engineering.  

5.2  Care instructions 

 

If dust is inside the integrating sphere, clean it with an air blower.  

 

Do not remove the fiber from the spectrometer. Otherwise, the calibration is invalid, and the 

spectrometer must be 

recalibrated

!  

5.3  Disposal instructions  

After the service life of the LE7, it must be disposed of properly. Electrical and electromechanical components 

are included in LE7. Observe your national regulations and ensure that third parties cannot use the LE7 after 

disposal.  

Contact Image Engineering if assistance for disposal is required. 

 

Summary of Contents for LE7 E

Page 1: ...ineering GmbH Co KG Im Gleisdreieck 5 50169 Kerpen Germany T 49 2273 99 99 1 0 F 49 2273 99 99 1 10 www image engineering com LE7 x2 x4 x6 E IR User manual for the LE7 product line x2 x4 x6 E IR January 14 2020 ...

Page 2: ...Overview display and ports 5 3 2 Connecting the hardware 7 4 OPERATING INSTRUCTIONS SOFTWARE 8 4 1 Requirements 8 4 2 Software installation 8 4 3 Starting the system 8 4 3 1 Spectrometer settings 8 4 3 2 Spectrometer calibration 8 4 3 3 iQ LED valibration 8 4 4 Low intensity use 8 4 4 1 Low light in combination with a neutral density chart 9 4 4 2 Using single iQ LED modules for low light 10 4 4 3...

Page 3: ...pectrometer is also included with the device Only suitable for indoor use Place your system in a dry constantly tempered environment without light interference The optimal ambient temperature range is 22 to 26 degrees Celsius The maximum ambient temperature range is 18 to 28 degrees Celsius The optimal system temperature range displayed in the software user interface is between 35 and 50 degrees C...

Page 4: ...ook directly into the sphere or light source when using high intensities or sequences with a low response time Do not open the device without instructions from the Image Engineering support team or when connected to the power supply 2 GETTING STARTED 2 1 Scope of delivery o Integrating sphere with a dual slot for D280 size charts o Spectrometer built in o Power cord o USB cable o Control software ...

Page 5: ... the first row of both iQ LED devices in the top and bottom borders of the image 5 From this position increase the distance of your camera to the LE7 so that both LED rows will move out of the image the minimum increase of distance should be 45 mm 6 Insert your test chart again and proceed with your test When using a diffuse test chart for shading or uniformity measurements you can move your camer...

Page 6: ... saved light sequence with different illuminants it is possible to save one sequence on the device Use the power button to turn on and off the light There are three pre stored illuminants on your device the intensity of each illuminant is shown in the acceptance protocol of your device 1 illuminant A default illuminant 2 illuminant D50 3 illuminant D75 Note To store your generated illuminants or s...

Page 7: ...ur test setup For example with an iQ Trigger see 2 1 optional equipment 3 2 Connecting the hardware Connect the power cord to the power outlet on your LE7 Connect the LE7 USB cable to your PC and turn on the LE7 the power switch is next to the power outlet Then connect the USB cable from the spectrometer calibration device to your LE7 The system will install the spectrometer and iQ LED driver on y...

Page 8: ...oftware see iQ LED software manual automatically generates the optimal spectrometer settings for your lighting conditions after pressing the auto detect button For special applications it is also possible to set the spectrometer settings manually If you have further questions please contact the Image Engineering support team 4 3 2 Spectrometer calibration Your spectrometer comes fully NIST traceab...

Page 9: ...r works correctly The spectrometer settings are correct All LED channels work correctly The dark measurement is correct Your measurement environment is correct Your ambient temperature is correct How to perform the spectral calibration is described in the iQ LED control software manual 4 4 Low Intensity use When using your system with very low intensity the spectral measurement values will start t...

Page 10: ...tion chart should correspond to an intensity of approximately 0 25 lx when using only the iQ LED module with the ND filter To get a rough idea of which intensities will appear when using only the ND module you can use a Density Filter Compensation value of 2 68 in the iQ LED SW Please see chapter 3 1 3 in the software manual for details Please note that it is impossible to measure such low intensi...

Page 11: ...mage Engineering Seite 11 von 11 6 TECHNICAL DATASHEET See annex for the technical data sheet It can also be downloaded from the website of Image Engineering https image engineering de support downloads ...

Reviews: