![Imada MV-500NII Series Instruction Manual Download Page 17](http://html1.mh-extra.com/html/imada/mv-500nii-series/mv-500nii-series_instruction-manual_2044913017.webp)
Instruction Manual MX-500N / MV-500NII Series
17
Manual No.
MXV50801E
Operation in Manual mode
1. Select Manual mode by
button.
2. Whenever pressing
toggles “FS Effective” LED lights up and off. When it is
light
s, “Speed knob” is not effective, but the “Fixed speed volume” is. You cannot
toggle FS Effective while the crosshead runs.
3. While pressing
the crosshead runs to the direction of button figure at its
maximum speed. It stops when you release the button or when it gets to each limit
knob.
4. While pressing
the crosshead runs to the direction of button figure at
the speed indicated by speed volume or FS volume. It stops when you release the
button or when it gets to each limit knob.
Operation in Single cycle mode
1. Select Single cycle mode by
button.
2. Whenever pressing
toggles “FS Effective” LED lights up and off. When it is
lights, “Speed knob” is not effective, but the “Fixed speed volume” is. You cannot
toggle FS Effective while the crosshead runs.
3. Press and release
buttons to start crosshead running to the direction of
button figure at its maximum speed. The crosshead reverses when it gets to the
limit knob at its maximum speed. It stops when it gets to another limit knob.
4. Press and release
buttons to start crosshead running to the direction of
button figure at the speed indicated by speed volume or FS volume. When it gets to
each limit knob, it goes reverse direction until getting to another limit knob.