PG 5
SAFETY
WARNING: Never use optical instruments such as microscopes,
eye loupes or magnifiers together with the original protective
eyewear. Otherwise sufficient eye protection can no longer be
ensured.
1.6.
Optical Fiber Cable and FastFiber Tip
Do not sharply bend, twist or knot the fiber cable, which leads to the
damage to the fiber and harm to the patient and laser operator.
Prior to each use, please examine the optical windows of the handpiece
body to make sure there are no dirt or dust contaminations which can
cause overheating and damage of the connector, raising the risk of fire
hazards.
If the optical cable or the fiber tip is damaged during operation, the
operator must immediately stop laser and get it replaced. Otherwise it
can cause overheat of the connector, raising the risk of fire hazards.
After the operation, please cover the distal end of the handpiece body
with the protective cap to keep clean of the optical windows.
The FastFiber tip is designed only for single use. It must be disposed of
after each use.