PG 4
SAFETY
WARNING: Use of controls or adjustments or performance of
procedures other than those specified herein may result in
HAZARDOUS radiation exposure.
WARNING: Do not aim the laser at metallic or reflective surfaces,
such as surgical instruments or dental mirrors. The laser beam
will be reflected to create potential hazards.
1.4.
Nominal Ocular Hazard Distance (NOHD)
The Nominal Ocular Hazard Distance (NOHD) of the iM3 Veterinary Dental
Laser is 3.7 meters @450nm, 0.9 meters @635nm, 4.4 meters @810nm,
and 5.1 meters @980nm, calculating from the distal end of the optical fiber.
1.5.
Laser Protective Eyewear
All persons present in the room must always wear the laser protective
eyewears from the iM3 Veterinary Dental Laser package.
Before using the laser protective eyewear, please make sure that your laser
protective eyewear:
are not damaged
conform to protection level L5 from standard EN 207
are suitable for the working wavelength (labeled on the eyewear)
These instructions apply particularly when using eyewear supplied from
outside sources that are not included in the scope of delivery of the
iM3
Veterinary Dental Laser
.