Fuse Rating 

Refer to the table below and check that the 
power supply voltage fits the fuse rating. To   
check the fuse rating, remove the fuse box. 
   
(2)

 

Filling Maintenance oil   
Check that the filter of the funnel is clean.   
Remove the oil cap and fill with the funnel.   
Fill it to the “full” level, while checking the   
oil level gauge. After filling oil, screw the oil 
  cap back on tight. 

 

DANGER 

 

Do not fill maintenance oil in the room with a risk of an explosion or near open flames. 

 

CAUTION 

Oil absorbent sheet 

Place an oil absorbent sheet at the bottom of the lubricating and cleaning room of the unit. The oil 
absorbent sheet absorbs the maintenance oil that was used for lubricating and cleaning. 

Connecting Power Cord   

Insert the power cord jack into the inlet at the back of the unit. 

 

 

Summary of Contents for HCR-100

Page 1: ...1 iM3 HCR 100 Automatic Handpiece Cleaning Lubrication System OPERATION MANUAL...

Page 2: ...UTIONS 3 1 FEATURES 5 2 SPECIFICATIONS 6 3 PARTS NAME 7 4 INSTALLATION AND ASSEMBLY 7 5 OPERATION PROCEDURES 8 6 DOOR 11 7 MAINTENANCE 11 8 TROUBLE SHOOTIING 12 SAFETY PRECAUTIONS Read these safety ca...

Page 3: ...use the product in direct sunlight or where the temperature exceeds 40C 104F Do not use the product for other purposes than the maintenance cleaning and lubrication of handpieces described in this op...

Page 4: ...sure above 0 35Mpa Do not open the oil cap during operation Open only after the unit completely stops When opening the oil cap in case of abnormal stop opening the door wait for approx 5 seconds befor...

Page 5: ...r effectively and efficiently cleans and lubricates contra angle handpiece Pressing the air key can purge excess oil from the handpiece after lubrication and cleaning cycle to some extent Select the l...

Page 6: ...6 2 SPECIFICATIONS Model SC L1 Rated Voltage AC120V10 AC230V10 50 60Hz Input Voltage 24 0VA Air Pressure 0 25 0 3Mpa 2 5 3 0kgf cm2 50 85psi Dimensions W280 D275 H360mm Weight 6 8kg...

Page 7: ...ll it to the full level while checking the oil level gauge After filling oil screw the oil cap back on tight DANGER Do not fill maintenance oil in the room with a risk of an explosion or near open fla...

Page 8: ...lose the door securely Contra Angle Handpiece 1 Remove the bur from the handpiece 2 Open the door of the unit and attach the handpiece to the E type joint until it locks 3 Point the bur insertion hole...

Page 9: ...cation cycle 5 When the cleaning and lubrication cycle is completed the mode indicator turns on and the buzzer sounds to tell the completion of the cycle 6 Pressing the air key blows out air only from...

Page 10: ...ng the air key 2 In 1 2 seconds or less press the start key The mode indicator changes at that time S L lighting The buzzer sounds L E lighting The buzzer does not sound 3 After selecting either press...

Page 11: ...leaning cycle 4 or 5 times before use to secure stable oil absorption of mist filters Alternatively apply maintenance oil directly to mist filters before use CAUTION Replace oil absorbent sheets appro...

Page 12: ...burned The unit does not clean and lubricate Does the mode indicator light Is the door open Remove the air tube press all three selection buttons and press start button Does the Unit produce an operat...