6A pH Sensor
An
evaluation
of the calibration process takes place here. If an error arises here then please repeat the
calibration process for this sensor.
6B Tank pH-value?
For the rH-value to be displayed the
pH-value in the tank
is required. Here you can either specify the pH-
value by selecting a pH sensor or by entering a value in the menu point "input". Confirm your selection with
the
enter
key.
6D Sensor in short-circuit plug
Connect the supplied short-circuit plug (it looks like a measurement module without an attached sensor lead)
to the measurement module
instead of the sensor
. Press the
enter
key.
6E Short-circuit plug
"Please wait" appears on the display. If this does not disappear after a few minutes then there is an error.
6F Sensor in Sol. 230 mV
The sensor must be re-connected to the measurement module. Then place the sensor in the 230V buffer
solution and press the
enter
key.
6G Sol. 230mV
"Please wait" appears on the display. If this does not disappear after a few minutes then there is an error. If
"
OK
" appears, please press the
enter
key.
6H Rx-Sensor
An
evaluation
of the calibration process takes place here. If an error arises here then please repeat the
calibration process. Press the
enter
key.
6K Sensor in Sol.2
Place the sensor in the
second buffer solution
.
6L Sol. 2
"Please wait" appears on the display. If this does not disappear after a few minutes then there is an error. If
"
OK
" appears, please press the
enter
key.
6Q Zero measurement
After two to three minutes “
Buffer 0.00
” should appear on the display. If this is not the case then the
measurement module may be defective.
6T Sensor OK
This display shows you that the
sensor is in working order
.