25
Error codes
13
Maintenance
14
The drive unit is maintenance-free but not immune from wear. The
carbon brushes of the motor wear down over the time.
For cleaning disconnect the main plug.
Only use cleansing agents which have been recommended by
IKA
.
Dirt
Cleaning agent
Dyes
isopropyl alcohol
Construction materials water containing tenside/isopropyl alcohol
Cosmetics
water containing tenside/isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA
application support.
Wear protective gloves during cleaning the instruments.
Electrical instruments may not be placed in the cleansing agent for
the purpose of cleaning.
Do not allow moisture to get into the instrument when cleaning.
Before using another than the recommended method for cleaning
or decontamination, the user must ascertain with
IKA
that this
method does not destroy the instrument.
Spare parts order:
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part, see
www.ika.com
,
spare parts diagram and spare parts list.
Turbo function will be stopped and rotation speed will return
to original setting. A symbol appears to indicate that the
Turbo function is stopped. Before this symbol disappears,
turbo function can not be activated again.
Any malfunction during operation may be identified by an error
message on the display.
Proceed as follows in such cases:
Switch off device using the power switch
Carry out corrective measures
Restart the device.
Error
code
Cause
Effect
Solution
Error 3
Inside temperature is
too high
motor off - Switch off the device and
let it cool down.
- Restart the device.
Error 4
Motor can’t run normal
(can’t start or not reach
the target speed)
motor off - Switch off the device.
- Check the structure for
possible blockades
Error 47
Error 48
Overload
motor off - Switch off the device
- Decrease the speed setting
or the load
Error 50
Tube is protected in
order to avoid overload
motor off - Check the tube condition
- Decrease the speed setting
or the load
If the actions described fails to resolve the fault or another error
code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Summary of Contents for ULTRA -TURRAX Tube Drive P control
Page 41: ...41 BMT 20 BMT 50 D IKA 40 C IKA 3...
Page 45: ...45 Power A Fig 1 9 A 6000 Timer 30 400 Timer Rev 10 Turbo 8000 Rev Timer Timer...
Page 50: ...50 16 IKA 24 15 www ika com IKA www ika com...
Page 65: ...65 IVD BMT 20 BMT 50 IKA 3 40 C IKA...
Page 73: ...73 13 14 IKA IKA IKA IKA www ika com Error 3 Error 4 Error 47 Error 48 Error 50...
Page 74: ...74 15 16 www ika com IKA www ika com IKA...
Page 91: ...91 IVD BMT 20 BMT 50 IKA 40 C IKA BG...
Page 93: ...93 IVD BMT 20 BMT 50 IKA 40 C IKA EL...
Page 94: ...94 Note...
Page 95: ......