background image

32

33

Commands

Function

IN_NAME

Read device name�

RESET

Switch to normal operation mode

IN_PV_3

Read external actual temperature

IN_PV_4

Read the actual speed (rpm)

IN_PV_5

Read the actual torque (M%)

IN_PV_80

Read visicocity

IN_PV_82

Read shear rate

IN_PV_83

Read shear stress

IN_PV_84

Read density

OUT_SP_4 X
(X = 0 ��� 200)

Adjust the rated speed value

OUT_SP_81 n

Select spindle

START_4

Start the motor

STOP_4

Stop the motor

IN_SOFTWARE_ID

Read software ID and version

/// Connection between the device and external devices:

PC 1�1 Cable: This cable is required to connect RS 232 port to a PC�

1
2 RxD

3 TxD

4

5 GND

6

7 RTS

8 CTS

9

1

RxD 2
TxD 3

4

GND 5

6

RTS 7

CTS 8

9

PC

1  

2  

3  

4  

5

6  

8  

9

9  

8  

7  

6

5  

4  

3  

2  

1

USB 2�0 Cable (A – micro B): This cable is required to connect USB port to a PC�

USB 2.0 micro B

USB 2.0 A

A

Error codes

The fault is shown by an error message on the display if the error occurs�
Proceed as follows in such case:
›  Switch the device off�
›  Carry out corrective measures�
›  Restart the device�

Error message

Cause

Effect

Correction

Error 1

cable or motor disc sensor 
damaged

motor disc no feedback signal

send back to service

Error 2

cable or spring disc sensor 
damaged

spring disc no feedback signal

send back to service

Error 4 

motor or motor driver 
damaged

device can't work

send back to service

If the action described fails to resolve the fault or another error code is displayed then take 
one of the following steps: 
›  Contact the service department 
›  Send the device for repair, including a short description of the fault� 

Maintenance and cleaning

The device is maintenance-free� It is only subject to the natural wear and tear of components 
and their statistical failure rate�

/// Cleaning:

›  For cleaning disconnect the mains plug!
›  Use only cleaning agents which have been approved by IKA to clean IKA devices�
  These are water (with tenside) and isopropanol�
›  Wear protective gloves during cleaning the device�
›  Electrical device may not be placed in the cleansing agent for the purpose of cleaning�
›  Do not allow moisture to get into the device when cleaning�
›  Before using another than the recommended method for cleaning or decontamination, the 

user must ascertain with IKA that this method does not destroy the device�

/// Spare parts order:

When ordering spare parts, please give:
›  device type,
› serial number, see type plate,
›  item and designation of the spare part see 

www.ika.com,

›  software version�

The above details correspond as far as possible to the recommendations of the NAMUR work-
ing party (NAMUR recommendations for the design of electrical plug connections for analogue 
and digital signal transmission on individual items of laboratory control equipment, rev� 1�1)�
The NAMUR commands and the additional specific IKA commands commissioning serve only 
as low level commands for communication between the device and the PC� With a suitable ter-
minal or communications program these commands can be transmitted directly to the device� 
The IKA software package, Labworldsoft

®

, provides a convenient tool for controlling device 

and collecting data under MS Windows, and includes graphical entry features, for motor speed 
ramps for example�

Summary of Contents for ROTAVISC

Page 1: ...IKA ROTAVISC...

Page 2: ...ENGLISH...

Page 3: ...on and leveling 17 Device self checking 18 Place the beaker and tempering bath in proper position 19 Select spindle 19 Assemble the selected spindle 20 Set the test speed 0 200 rpm 22 Start the test...

Page 4: ...ained staff work with the device Follow the safety instructions guidelines occupational health safety and accident prevention regulations The device must only be used in a technically perfect conditio...

Page 5: ...isposal instructions The device packaging accessories and batteries must be disposed of in accordance with national regulations Use The IKA ROTAVISC must be used in conjunction with the IKA ROTASTAND...

Page 6: ...O P Q R S M ROTASTAND pillar N ROTASTAND base O Screw washer P Hexagon socket screw M8 Q Hexagon socket screw M4 R Screwdriver AF 2 5 mm DIN ISO 2936 S Screwdriver AF 5 mm DIN ISO 2936 Assembly 1 2 1...

Page 7: ...1 2 1 2 Click Push 8 9 24V 100 mA LPS 1 2 3 Note The protection cap protects the coupling nut and the inside pivot from damage during transportation Keep the protection cap in the hole of the stand in...

Page 8: ...button Return to previous screen Menu button Touch it once Enter main menu screen Touch it a second time Back to working screen Control knob Rotating Pressing Start Stop the motor for measuring press...

Page 9: ...symbols on the working screen Symbol Designation Function Profile symbol Indicate the selected profile Accurate mode symbol Indicate the accurate mode is selected Balanced mode symbol Indicate the ba...

Page 10: ...pa s Accuracy 22 mpa s Spindle SP 4 Speed 28 20 rpm Maximum Viscosity 226950 mpa s Accuracy 2269 mpa s ON O O ON N N N O ON O ON O O ON N N N O ON O ON O O ON N N N O ON O 2 ON O O ON N N N O ON O 1 2...

Page 11: ...ity Spindle Search Spindle SP 1 Speed 28 20 rpm Maximum Viscosity 2 269 mpa s Accuracy 22 mpa s None Profile 1 SP 1 Stop 0 mPa s rpm n 0 00 00 00 00 0 0 M y Accuracy y Viscosity Spindle Search Spindle...

Page 12: ...dit the values or settings or activate deactivate a function Rotate the control knob to OK or press the Back button or Menu button to end the procedure and return to the previous menu Note The menu op...

Page 13: ...Visc Display Firmware Update Screen Background Black White 100 Sound Volume Factory Settings Communication Device Name ROTAVISC Information Viscosity Software Control mode Density Version Spindle 1 0...

Page 14: ...ram Program In this menu 5 user defined speed time profiles can be created A program can consist of up to 10 segments Once a program has been selected the following options are available 1 Start Start...

Page 15: ...e the control mode as Accurate Balance or Fast For details see Accu racy under menu option Viscosity 2 Start The user can define the start speed 0 01 to 200 rpm for spindle 3 End The user can define t...

Page 16: ...ika com ika lws download usb driver zip Install the driver by running the setup file Then connect the IKA device through the USB data cable to the PC The data communication is via a virtual COM port R...

Page 17: ...the device for repair including a short description of the fault Maintenance and cleaning The device is maintenance free It is only subject to the natural wear and tear of components and their statis...

Page 18: ...ion Connection for external temperature sensor PT 100 Working temperature C 100 300 Working temperature display TFT LCD Temperature measurement resolution K 0 1 Temperature measurement accuracy 100 C...

Page 19: ...s Spindle lo vi me vi hi vi l hi vi II SP 1 6 000 64 000 128 000 512 000 SP 2 30 000 320 000 640 000 2 560 000 SP 3 120 000 1 280 000 2 560 000 10 240 000 SP 4 600 000 6 400 000 12 800 000 51 200 000...

Page 20: ...Materials Determination of Viscosity of Crude or Modified Isocyanates D 5018 89 Standard Test Method for Shear Viscosity of Coal Tar and Petroleum Pitches Reapproved 1999 D 5133 01 Standard Test Meth...

Page 21: ...00 eMail info ika in UNITED KINGDOM IKA England LTD Phone 44 1865 986 162 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangz...

Reviews: